Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
207
SOUR:POW:SLOP
SOURce<Ch>:POWer[:LEVel]:SLOPe[:DATA]
<power>
SOURce<Ch>:POWer[:LEVel]:SLOPe[:DATA]?
Description
Sets
or
reads
out
the
power
slope
value
for
the
frequency
sweep
(command/query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<
power
>
the
power
slope
value
from
–2
to
+2
Resolution
0.1
Unit
dB/GHz
(decibel/gigahertz)
Out
of
Range
Sets
the
value
of
the
limit,
which
is
closer
to
the
specified
value.
Query
Response
<numeric>
Preset
Value
0
Equivalent
Softkeys
Stimulus
>
Power
>
Slope
[dB/GHz]