Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
149
SENS:CORR:COLL:CKIT
SENSe<Ch>:CORRection:COLLect:CKIT[:SELect]
<numeric>
SENSe<Ch>:CORRection:COLLect:CKIT[:SELect]?
Description
Sets
or
reads
out
the
number
of
the
selected
calibration
kit
(command/query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<numeric>
the
number
of
the
calibration
kit
from
1
to
11
Out
of
Range
Error
occurs.
The
command
is
ignored.
Error
code:
222.
Query
Response
<numeric>
Preset
Value
1
Equivalent
Softkeys
Calibration
>
Cal
Kit
>
Cal
Kit
n
SENS:CORR:COLL:CKIT:LAB
SENSe<Ch>:CORRection:COLLect:CKIT:LABel
<string>
SENSe<Ch>:CORRection:COLLect:CKIT:LABel?
Description
Sets
or
reads
out
the
calibration
kit
label
(command/query).
Target
Calibration
kit,
selected
for
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<
string
>,
up
to
254
characters
Query
Response
<
string
>
Preset
Value
Varies
depending
on
the
number
of
the
calibration
kit.
Equivalent
Softkeys
Calibration
>
Edit
Cal
Kit
>
Label