57
CALC:LIM:REP:ALL?
CALCulate<Ch>[:SELected]:LIMit:REPort:ALL?
Description
Reads
out
the
data
array,
which
is
the
limit
test
results
(query
only).
The
array
size
is
4N,
where
N
is
the
number
of
measurement
points.
For
the
n–th
point,
where
n
from
1
to
N:
<numeric
4n–3>
the
stimulus
value
in
the
n–th
point;
<numeric
4n–2>
the
limit
test
result
in
the
n–th
point;
–1:
No
limit
0:
Fail
1:
Pass
<numeric
4n–1>
the
upper
limit
value
in
the
n–th
point
(0
–
if
there
is
no
limit)
<numeric
4n–0>
the
lower
limit
value
in
the
n–th
point
(0
–
if
there
is
no
limit)