Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
169
SENS:CORR:COLL:METH:TYPE?
SENSe<Ch>:CORRection:COLLect:METHod:TYPE?
Description
Reads
out
the
calibration
type
selected
for
the
calculation
of
the
calibration
coefficients
on
completion
of
the
calibration
executed
by
the
SENS:CORR:COLL:SAVE
command
(query
only).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Query
Response
"RESPO"
:
Response
(Open)
"RESPS"
:
Response
(Short)
"RESPT"
:
Response
(Thru)
"SOLT1"
:
Full
1–port
calibration
"1PATH"
:
One
path
2–port
calibration
"NONE"
:
Not
defined
Equivalent
Softkeys
None
SENS:CORR:COLL:OPEN
SENSe<Ch>:CORRection:COLLect[:ACQuire]:OPEN
<port>
Description
Measures
the
calibration
data
of
the
open
standard
for
the
specified
port
(no
query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<port>
the
number
of
the
port
from
1
to
2
Out
of
Range
Error
occurs.
The
command
is
ignored.
Error
code:
222.
Notes
The
command
start
the
measurement
for
the
channel
independently
of
the
trigger
and
trigger
source
settings.
The
command
waits
for
the
completion
of
the
measurement.
Equivalent
Softkeys
Calibration
>
Calibrate
>
Response
(Open)
>
Open
Calibration
>
Calibrate
>
Full
1–Port
Cal
>
Open
Calibration
>
Calibrate
>
One
Path
2–Port
Cal
>
Open