Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
185
SENS:FREQ:STAR
SENSe<Ch>:FREQuency:STARt
<frequency>
SENSe<Ch>:FREQuency:STARt?
Description
Sets
or
reads
out
the
stimulus
start
value
of
the
sweep
range
for
linear
or
logarithmic
sweep
type
(command/query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<
frequency
>
the
stimulus
start
value
from
3E5
to
1.3E9
(Planar
–
TR1300/1),
from
2E4
to
4.8E9
(TR5048),
from
2E4
to
3.0E9
(TR7530).
Unit
Hz
(Hertz)
Out
of
Range
Sets
the
value
of
the
limit,
which
is
closer
to
the
specified
value.
Query
Response
<numeric>
Preset
Value
3E5
(Planar
–
TR1300/1),
2E4
(TR5048),
2E4
(TR7530).
Equivalent
Softkeys
Stimulus
>
Start