Introduction to MegaPulse 

 

The  impulse  test  is  designed  to  simulate  impulse  surges  which  occur  in  everyday  life  due  to 
nearby  lightning  strikes,  switching  transients,  and  other  high-frequency  faults  on  the  power 
distribution network.  Impulse testing is the fundamental method for empirical verification of the 
adequacy  of  insulation.  Other  methods  of  ensuring  adequate  insulation  (AC  or  DC  Dielectric 
Withstand  testing,  measurement  of  over-surface  creep  age,  through-air  clearance,  or  distance-
through-insulation)  are  all  extrapolated  from  the  results  of  impulse  testing.    The  impulse  test  is 
performed  to  ensure  that  the  insulation  in  question  will  be  able  to  function  properly  when 
subjected to similar impulse surges in the field.  

 

Safety Precautions 

 

The  impulse  withstand  test  can  generate  voltages  up  to  5000V  peak  at  potentially  lethal  current 
levels.    Currents  of  as  little  as  5mA  at  120  volts  can  cause  death;  the  MegaPulse  can  deliver 
currents of more than 10 Amps peak for very short time duration. The potential for serious injury 
or death exists and personnel should be aware when they conduct this test. 

 

Test Personnel 

 

Personnel require special training to conduct the impulse test.  They should understand electrical 
fundamentals clearly, and be aware that high voltage is adept and creative at completing a path to 
ground.    Instructions  should  include  a  warning  against  any  metal  jewelry.    Operators  should  not 
allow others in the testing area, especially when tests are being conducted.  Organization is to be 
stressed.  The operator should keep the area free of unused leads and equipment. 
 

Testing Area 

 

The  area  used  for  conducting  the  impulse  test  should  be  as  remote  as  possible  from  normal 
production  line  activities.    Only  personnel  actually  conducting  the  test  should  be  allowed  in  the 
area, and it should be taped or roped off to preclude casual entry by other employees.  In addition, 
the area should be marked "WARNING - HIGH VOLTAGE TESTING" or the equivalent to warn 
others of the nature of the testing taking place. 
 
The  bench  being  used  should  be  non-conductive,  and  any  exposed  metal  parts  should  be  tied 
together and grounded.  If a conductive surface must be used, it should be grounded. Because of 
sparking  during  an  impulse  test  failure,  it  is  not  safe  to  conduct  impulse  testing  in  combustible 
atmospheres. 
 
It  is  imperative  that  a  good  ground  be  provided  to  the  MegaPulse  tester.    Before  connecting  the 
equipment,  ensure  that  the  building  wiring  provides  a  low-resistance  ground.    If  the  MegaPulse 
tester is used on a high-resistance grounding circuit, dangerous high voltages may be present to the 
operator.    In  addition,  the  power  to  the  Testing  Area  should  be  provided  with  an  easily  reached 
shutoff switch which can be actuated by personnel outside the Area if needed. 
 

 
 

Summary of Contents for MegaPulse Defib-5PF-002

Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ......

Page 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Page 5: ...5...

Page 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Page 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Page 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Page 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Page 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Page 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Page 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Page 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Page 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Page 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Page 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Page 21: ...21 Appendix 1 Commands table...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Reviews: