17
Figure 7. Waveform Measurement Setup
Testing
This section describes how the MegaPulse Defib-5PF-002 tester is used to conduct a test. The test
can be stopped immediately at any time by turning OFF the rear-panel power switch.
1.
Select the input voltage (115V or 230V) from the switch on the back and connect the tester
to a proper source of supply using the included 18AWG power supply cord.
2. Plug the leads to the Outputs of the tester
3.
Connect the ends of the test leads to the equipment under test.
4.
Connect the Computer to the TestMinder port on the back of the tester
5.
Turn the Tester on.
5.
Use table of commands to operate the tester.
6.
Set desired voltage.
7.
Select multiplexor channel.
8.
Charge.
9.
Trigger
Note: Polarity is disable and the multiplexor can only be operated via commands.