Initial Checkout Procedure 

 

The  following  procedure  will  verify  that  the  MegaPulse  Defib-5PF-002  tester  is  working 
correctly.  We  recommend  that  this  procedure  be  conducted  periodically  to  ensure  proper 
operation  of  the  tester.    The  following  items  are  needed  to conduct this procedure: A measuring 
instrument to monitor the output waveform. Always ensure that the measuring instrument (usually 
an oscilloscope with a high-voltage probe) is rated for the voltage involved, and that the frequency 
response  of  the  instrument  and  probe  is  capable  of  measuring  the  output  waveform  of  the 
MegaPulse  Defib-5PF-002  tester.    A  measuring  instrument  or  probe  with  a  low  frequency 
response will result in erroneous readings that could be mis-read.  

  

CAUTION 

High voltage generated by the MegaPulse tester is exposed during this test. A risk of 
shock exists.  Exercise care when using the MegaPulse tester. 

 

1. 

Connect the tester to a proper source of supply using the included 18 AWG power supply  

2. 

Plug a pair of leads to the Output on the front panel. Use channel 1 and 2 for this test. 

3. 

Connect  the  leads  to  an  appropriate  measuring  instrument  (typically  an  oscilloscope  with  a 

high-voltage probe). Note that the 

RETURN 

lead is referenced to the chassis ground of the 

tester, in this case we will connect channel 1 to the return of the probe, connect channel 2 to 
the High Voltage of the probe.  

4. 

 Turn  the  Tester  on.  And  connect  the  computer  to  the  TestMinder  port  on the rear panel of 

the tester. 

5. 

For the following steps refer to the commands table. 

 

5.1 Send command to select channel 2 output 

 

5.2 Send command to set voltage at 500V 

 

5.3 Send command to charge 

 

5.4 Send command to trigger 

6. 

Capture de waveform on the oscilloscope. 

 

Note that the peak amplitude of the measured output waveform is proportional to the voltage 

that is read of the front panel of the MegaPulse, but it will always be somewhat lower.  This 
is because the meter on the MegaPulse is measuring the voltage on the main impulse storage 
capacitor.  This voltage will intentionally dissipate to some extent before reaching the output 
leads. 

7. 

Repeat step 5 charging at 5000V. 

8. 

Manual control checkout. Note: Multiplexor control can only be set via commands 

 

8.1 Turn the VOLTAGE knob to 500V and press the knob to set the voltage 

 

8.2 Press the charge button, verify the meter reach the set voltage. 

 

8.3 Press TRIGGER to discharge. 

9. 

Capture the waveform on the oscilloscope. 

10  Repeat step 8 at 5000V. 

 

 

 

 

 

Summary of Contents for MegaPulse Defib-5PF-002

Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ......

Page 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Page 5: ...5...

Page 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Page 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Page 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Page 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Page 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Page 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Page 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Page 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Page 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Page 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Page 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Page 21: ...21 Appendix 1 Commands table...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Reviews: