14
Table 1. MegaPulse Defib-5PF-002 Front Panel
ITEM
NAME
FUNCTION
1
VOLTAGE
Adjust Knob
Adjust the digital voltage set point in the tester.
Press the voltage knob to display the voltage set point. This setting will blink for a
few seconds on the front meter.
Turn Clockwise to increase the setting Voltage Setting Point before pressing
CHARGE
button.
2
POLARITY
switch
Disable
3
NOR REV
indicator
Indicates the state of the Output Polarity switch.
NOR
indicates Normal (Positive) position.
REV
indicates Reverse (Negative) position.
4
CHARGE
switch
Starts the charge process of the tester capacitor.
The
CHARGE
indicator will turn off after the
CHARGE
switch is pressed, and the
TRIGGER
indicator will turn on.
The charge process will stop after 2 minutes if the
TRIGGER
button is not pressed.
5
CHARGE
indicator
This Yellow indicator is lit to show that pressing the
CHARGE
switch is the next
logical step in a test sequence.
CHARGE
indicator is lit when the tester is turn ON an after pressing
TRIGGER
button.
CHARGE
indicator will go out after pressing
CHARGE
button.
CHARGE
and
TRIGGER
Indicators will be blinking if the Interlock Switch is open.
(Only testers with Interlock Switch Option)
6
TRIGGER
switch
Triggers the output impulse waveform.
The impulse waveform will appear across the output leads.
7
TRIGGER
indicator
This Red indicator is lit to show that the tester can be trigger.
TRIGGER
indicator is lit for 2 minutes after the
CHARGE
button is pressed.
TRIGGER
indicator will go out after pressing
TRIGGER
button.
TRIGGER
and
CHARGE
Indicators will be blinking if the Interlock Switch is open
(Only testers with Interlock Switch Option)
TRIGGER
indicator will blink at when the Voltage. This effect will remain on until
the
TRIGGER
switch is pressed. (Only testers with PC Interface option)
8
VOLTAGE
meter
Displays the output voltage set point.
The voltage reading will increase from zero to the voltage set point when the
CHARGE button is pressed. Note that the Voltage meter may indicate that some
residual voltage is present on the main storage capacitor, even when the tester is first
turned ON. This is due to inherent charging of the internal capacitors. Pressing the
TRIGGER
switch will discharge the capacitors. Note that the peak amplitude of the
measured output waveform is proportional to the voltage that is read of the front
panel of the MegaPulse, but it will always be somewhat lower. This is because the
meter on the MegaPulse is measuring the voltage on the main impulse storage
capacitor. This voltage will intentionally dissipate to some extent before reaching the
output leads.
The meter will start to flash at
7100V
to indicate that voltage is in the maximum
limits.
If unit includes PC Interface and the Keyboard is locked, the display will show OFF
when a button is pressed.
9
500:1
BNC
500:1 reference to verify delivery of pulse
10-19
OUTPUTS
10 Channel multiplexer outputs, each channel capable of delivering the pulse, channel
activation is via commands on the back of the unit.
Summary of Contents for MegaPulse Defib-5PF-002
Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...
Page 3: ......
Page 5: ...5...
Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...
Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...
Page 21: ...21 Appendix 1 Commands table...
Page 22: ...22...
Page 23: ...23...
Page 24: ...24...