background image

 

Section 2

 

 

Introduction 

 

This  manual  contains  complete  operating,  maintenance  and  calibration  information  for  the 
Compliance West USA MegaPulse Defib-5PF-002 Impulse Tester.   
 

 

In  case  of  trouble,  the  test  can  be  immediately  terminated  at  any  time  by  turning  the  rear-   

panel power switch to the OFF position.   

 

Before  the  test  can  commence, the unit must be armed by pressing the 

CHARGE

 Button.  

The test will not begin until the 

TRIGGER

 Button is pushed.   

 

Operator instructions are printed on the rear panel for quick reference.   

 

Voltage is discharged by a resistor bank within the MegaPulse tester upon test completion.  
Discharge progress is shown on the front panel meter. 

 
 
Your tester is warranted for a period of one year upon shipment of the instrument to the original 
purchaser.  
 
 

Specifications 

 

Specifications for the MegaPulse Defib-5PF-002 are listed in the table below and the component 
designations are shown in Figure 1 and Figure 2. For impulse waveform look at  
Figure

 3

3 and Figure 4. 

 

 

ELECTRICAL 

 
Output Voltage: 
 

0 - 5000 V tolerance ±1% 

Main Capacitance: 
 

0.13 uF ±10% Type A, 0.50 uF ±10% Type B (600,000 cycles life) 

Voltage Control: 
 

Digital  Set  point  adjusted  by  frontal  VOLTAGE  knob  or  by  PC 
TestMinder. 

Voltage Display: 
 

4 Digit LED Display. 

Voltage Meter resolution: 
 

2V 

Duty Cycle: 
 

1 pulse every 20s. 

Selectable Input Voltage: 

120V~ 60Hz 
240V~ 50Hz 

 

 

ENVIROMENTAL 

 

 

Operating Temperature: 

15 - 40 ºC 

Relative Humidity Range: 
Altitude 

0 - 90% non-condensing 
0 - 3000 meters 

 

 

GENERAL 

 

 

Dimensions: 

17” wide x 15.75” high x 17” in deep 

Weight: 

50 lbs. approx. 

 

 

Summary of Contents for MegaPulse Defib-5PF-002

Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ......

Page 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Page 5: ...5...

Page 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Page 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Page 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Page 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Page 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Page 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Page 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Page 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Page 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Page 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Page 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Page 21: ...21 Appendix 1 Commands table...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Reviews: