background image

 

Safety Techniques 

 

The high voltage circuit of the MegaPulse Defib-5PF-002 can be shut off at any time by turning 
OFF the rear power switch.  Always press TRIGGER to discharge the tester before turning OFF. 

 

The  MegaPulse  tester  is  provided  with  a  digital 

VOLTAGE  ADJUST

  knob  on  the front panel.  

This  should  always  be  adjusted  to  the  minimum  position  at  the start of testing.  In addition, this 
knob should be adjusted back to the minimum position at the completion of all testing. 

 

The  MegaPulse  tester  is  provided  with  a 

CHARGE

  switch  that  is  in  the  unarmed  "Standby" 

setting when the tester is first turned ON.  When the yellow 

CHARGE

 button is lit, the tester will 

not  provide  high  voltage  until  the 

CHARGE

  Button  and  the 

TRIGGER

  Button  have  been 

pressed in order.  To prevent inadvertent operation, the operator should be instructed not to press 
the 

CHARGE

 Button until the test is ready. 

 
The MegaPulse tester has been designed for one-touch operation with the right hand.  If possible, 
it should be set up to the left and in front of the equipment under test.  The equipment under test 
should  be  connected  to  the  MegaPulse  tester  and  then  left  alone  by  the  operator.    After  the 
operator is clear of the Tester and the equipment under test, the operator should turn the rear-panel 
power  switch  to  ON,  press  the 

CHARGE

  Button,  adjust  the  voltage  to  the  desired  level  (as 

displayed on the front panel meter), then press the 

TRIGGER

 Button, with his right hand.  This 

will allow the greatest separation between the operator and the test being conducted. Note that the 
CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed. 
 
The MegaPulse tester is designed to bleed the high voltage away after the test has concluded.  In 
order  to  ensure  that  any  voltage  present  in  the  equipment  being  tested  has  been  completely bled 
away, the operator should not unplug the equipment under test from the MegaPulse until the front 
panel meter reads a safe level (40 volts or less is generally considered a safe level).  Pressing the 
TRIGGER button before disconnecting main power (or turning the equipment off) will ensure that 
the internal capacitors are discharged as much as possible. 
 

Using the MegaPulse Impulse Tester 

 
The impulse test involves high voltage and caution should be exercised when using the tester.  The 

RETURN

  lead  is  referenced  to  building  ground  when  properly  connected.    However,  both  the 

OUTPUT

 and 

RETURN

 leads must always be treated as Hazardous whenever the power switch 

of the MegaPulse is in the ON position.  
 
The MegaPulse impulse tester generates the impulse waveform only; it does not determine Passing 
or Failing results.  It is the operator’s responsibility to monitor the output waveform and determine 
Passing  or  Failing  results.    In  monitoring  the  impulse  waveform,  consider  the  following  points: 
The  Impulse  waveform  is  high  voltage  and  high  frequency  (short  duration).    Always  ensure  that 
the  measuring  instrument  (usually  an  oscilloscope  with  a  high-voltage  probe)  is  rated  for  the 
voltage  involved,  and  that  the  frequency  response  of  the  instrument  and  probe  are  capable  of 
measuring  the  output  waveform  of  the  MegaPulse  Impulse  Tester.    A  measuring  instrument  or 
probe with a low frequency response will result in erroneous readings that could be mis-read. 

  

 
 
 

 

Summary of Contents for MegaPulse Defib-5PF-002

Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ......

Page 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Page 5: ...5...

Page 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Page 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Page 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Page 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Page 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Page 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Page 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Page 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Page 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Page 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Page 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Page 21: ...21 Appendix 1 Commands table...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Reviews: