Section 3

 

Operation 

 

This section describes how to set up and make measurements with your Tester.  We recommend 
that you read the entire section carefully so that you can use all the features of your Tester. 

 

Setting up your Tester 

 

Your  Tester  is  shipped  in  a  special  protective  container  that  should  prevent  damage  to  the 
instrument  during  shipping.    Check  the  shipping  order  against  the  contents  of  the  container  and 
report any damage or short shipment to Compliance West USA.  The container should include the 
following: 
 

 

Qty 1  MegaPulse Defib-5PF-002 Tester 

 

Qty 10 PN:00-HVTL-R (High Voltage Test Lead Red) 

 

Qty 1   PN: 70-101 (Power Cord 18 AWG) 

 

Qty 1   Defib-5PF-002 Instruction Manual 

 

Qty 1 Testminder set 

 

Qty 1 External Load (52.299kΩ and 7.492kΩ +/-5%) 
 

If  reshipment  of  the  instrument  is  necessary,  please  use  the  original  shipping  container.    If  the 
original shipping container is not available, be sure that adequate protection is provided to prevent 
damage  during  shipment.    We  recommend  that  the  instrument  be  surrounded  by  at  least  three 
inches of shock-absorbing material on all sides of the container. 
 
Remove the tester from its container and place it on a test bench.

   

 

AC Line Voltage Requirements 

 

AC  line  voltage  requirements  for  your  Tester  are  noted  on  the  rear  panel  of  the  instrument.  Do 
not connect the instrument to a different voltage source.  The cord packaged with your MegaPulse 
Tester is for use in the United States.  If another power cord must be used, the cord must be rated 
for the maximum current noted on the rear panel.  It must also meet the requirements of IEC 227 
or IEC 245, and mains cords that are certified or approved by any recognized national test house 
are regarded as meeting this requirement.   

Fuse Replacement 

 

There  is  a  user-replaceable  fuse  (F1)  located  on  the  rear  panel  of  the  instrument.    It  is  located 
behind a door in the Power Inlet-Power Switch-Fuse Holder device.  The fuse rating is noted on 
the rear panel.  Do not attempt to replace it with a fuse of any other rating. A second fuse can be 
found next to the input voltage selector switch 
 
Use the following procedure to replace a fuse: 
1. 

Turn the power switch to the OFF position. 

2. 

Unplug the instrument from the source of supply. 

3. 

Remove the power inlet cord from the instrument. 

4. 

Open the fuse holder door. 

5. 

Replace the fuse with a new one of the correct rating. 

6. 

Replace the fuse holder and power inlet cord.

   

Summary of Contents for MegaPulse Defib-5PF-002

Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Page 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 3: ......

Page 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Page 5: ...5...

Page 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Page 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Page 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Page 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Page 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Page 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Page 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Page 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Page 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Page 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Page 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Page 21: ...21 Appendix 1 Commands table...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Reviews: