12
Section 3
Operation
This section describes how to set up and make measurements with your Tester. We recommend
that you read the entire section carefully so that you can use all the features of your Tester.
Setting up your Tester
Your Tester is shipped in a special protective container that should prevent damage to the
instrument during shipping. Check the shipping order against the contents of the container and
report any damage or short shipment to Compliance West USA. The container should include the
following:
•
Qty 1 MegaPulse Defib-5PF-002 Tester
•
Qty 10 PN:00-HVTL-R (High Voltage Test Lead Red)
•
Qty 1 PN: 70-101 (Power Cord 18 AWG)
•
Qty 1 Defib-5PF-002 Instruction Manual
•
Qty 1 Testminder set
•
Qty 1 External Load (52.299kΩ and 7.492kΩ +/-5%)
If reshipment of the instrument is necessary, please use the original shipping container. If the
original shipping container is not available, be sure that adequate protection is provided to prevent
damage during shipment. We recommend that the instrument be surrounded by at least three
inches of shock-absorbing material on all sides of the container.
Remove the tester from its container and place it on a test bench.
AC Line Voltage Requirements
AC line voltage requirements for your Tester are noted on the rear panel of the instrument. Do
not connect the instrument to a different voltage source. The cord packaged with your MegaPulse
Tester is for use in the United States. If another power cord must be used, the cord must be rated
for the maximum current noted on the rear panel. It must also meet the requirements of IEC 227
or IEC 245, and mains cords that are certified or approved by any recognized national test house
are regarded as meeting this requirement.
Fuse Replacement
There is a user-replaceable fuse (F1) located on the rear panel of the instrument. It is located
behind a door in the Power Inlet-Power Switch-Fuse Holder device. The fuse rating is noted on
the rear panel. Do not attempt to replace it with a fuse of any other rating. A second fuse can be
found next to the input voltage selector switch
Use the following procedure to replace a fuse:
1.
Turn the power switch to the OFF position.
2.
Unplug the instrument from the source of supply.
3.
Remove the power inlet cord from the instrument.
4.
Open the fuse holder door.
5.
Replace the fuse with a new one of the correct rating.
6.
Replace the fuse holder and power inlet cord.
Summary of Contents for MegaPulse Defib-5PF-002
Page 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...
Page 3: ......
Page 5: ...5...
Page 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...
Page 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...
Page 21: ...21 Appendix 1 Commands table...
Page 22: ...22...
Page 23: ...23...
Page 24: ...24...