23
Section 6
Options
This section contains a list of available options for D5-PF.
Option Name
Description
34
Incorporates a circuit for Invasive Blood Pressure test (IEC 60601-2-34 Issue 2)
100X
HV jumper to disconnect the internal 100ohm resistor and use external Energy
measurement devices like Fluke 7000DP and 7010.
MPxxx
Different Line voltages, replace on the part number xxx for 100, 110, 220, 230 or 240
depending the line voltage. (120V for standard configuration).
SG
Built in Sine Wave Generator 22Vp-p (Max).
GVAC
Vacuum trigger relay (20sec. duty cycle only)
Circuit for Invasive Blood Pressure test (IEC 60601-2-34 Issue 2)
This option gives you the convenience to select the discharge resistor 100ohm or 100kohm with a high voltage jumper
located in the front panel; it also has an additional output impedance of 50kohm and switch to select trigger duration
of 200ms or 20 seconds, this resistance values are need when testing to IEC 60601-2-34. Ed. 2.0, If testing IEC 60601-
2-34. Ed. 3.0 no need of adding option 34.
100X
The 100X Option is offered as a separate upgrade to the unit. The Purpose of the 100X Option is to disconnect the
internal 100Ω resistor and allow the user to connect an external load for energy measurement purposes.
To connect or disconnect the internal 100 resistor, the D5-PF with option 100X is equipped with two high voltage
red connectors located on the rear panel. These two connectors are labeled with:
SHORT FOR INTERNAL LOAD:
Internal 100Ω load connected.
OPEN FOR EXTERNAL LOAD:
Internal 100Ω is disconnected (Note: Use an external load at the
Energy Measurement and Ground connectors).
WARNING:
If the Jumper is OPEN FOR EXTERNAL LOAD and the external load is not connected the unit will not discharge
properly. The external load should not be less than 95Ω or more than 1000Ω.
\
Summary of Contents for MegaPulse D5-PF
Page 1: ......
Page 2: ......
Page 5: ...iii Cleaning 26 Calibration Information 26 Technical Assistance 27...
Page 6: ...iv...
Page 16: ...10 Figure 2 Controls Indicators Connectors MegaPulse D5 PF Rear Panel...
Page 23: ...17 Figure 7 Y1 Y2 Test Output...