ITEM

 

NAME

 

FUNCTION

 

11 

400 ohm OUTPUT

 jack 

The impulse waveform appears on the 

OUTPUT

 jack, referenced to the 

RETURN 

jack. 

When the 

POLARITY

 switch is in the Normal position (

NOR

 indicator is lit) the output will be a positive pulse. 

When the 

POLARITY

 switch is in the Reverse position (

REV

 indicator is lit) the output will be a negative pulse. 

12 

AUX

 

OUTPUT 

jack 

The 

AUX OUTPUT 

is connected at the grounded side of the 100ohm resistor, but this output is only used when it is required to 

float the 100ohm resistor. 

13 

MEAS. POINT

 jack 

Used to perform the Pass/Fail reference test. 

14 

Y1 – Y2

 outputs 

These differential outputs are to be connected to two different channels of an oscilloscope.  The outputs are then subtracted and 
the resulting waveform is used to judge acceptability of the EUT.  

15 

ENERGY MEAS.

 switch

 

By pressing this switch, the front display will blink showing for a few seconds the energy measurement from the last pulse. 
Note: This function only works when the MegaPulse unit is not charging or triggering. 

16 

PASS FAIL REF.

 switch 

 

The D5-PF Includes the Pass-Fail reference circuit built in, to use properly refer to the directions on front panel. 

17 

PASS FAIL REF.

 indicator 

This Green indicator is lit to show that the pass/fail reference is connected (AUX OUTPUT must be grounded). 

18 

AUX OUTPUT 

switch 

It is used to control the AUX output position (Floated or Grounded). 

19 

AUX OUTPUT 

indicator 

This Green indicator is lit to show that 

AUX OUPUT

 switch is set on the 

FLOAT

 position. 

20 

S3 

switch 

Used for common mode testing of figure 201.103, S3 switch should be closed only when auxiliary output is floated from ground. 

21 

S3 

indicator

 

This Green indicator is lit to show that 

S3

 switch is set on the 

SHORT

 position. 

22 

INDUCTOR 

left switch 

This switch allows the user to select the inductance value between 500uH and 25mH.

 

23 

INDUCTOR 

right switch 

This switch allows the user to select the inductance resistance value between 10 ohm and 11ohm when INDUCTOR left switch is 
on 25mH.

 

24 

S2 

left

 

switch 

This switch is provided to protect the Signal Generator from the output pulse of the D5-PF when OFF, and to allow the required 
20V p-p signal to be injected when SINE WAVE is selected.  

25 

S2 

right

 

switch 

This switch is enable only when S2 left switch is on SINE WAVE position and allows selecting between 390K or 470K ohm.  

26 

BNC INPUT 10Hz, 22Vp-p

 

BNC Input provided on front panel is for connecting the output signal generator.  See Figure 12 Part 9A. 

Note:

 If the unit includes the Sine Wave Generation option, the item 26 will be a BNC OUTPUT 10Hz, for monitoring the Sine 

Wave with an oscilloscope. 

27 

COARSE

 adjustment (optional)

 

It is used to adjust the amplitude of the sine waveform. Only for models with Sine Wave Generation option. 

28 

FINE

 adjustment (optional)

 

It is used to fine adjust the amplitude of the sine waveform. Only for models with Sine Wave Generation option. 

 

Table 1.  Controls, Indicators, Connectors – MegaPulse D5-PF Front Panel 

Summary of Contents for MegaPulse D5-PF

Page 1: ......

Page 2: ......

Page 3: ...nd ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this man...

Page 4: ...eyboard and Voltage Knob Enable 13 Discharging Residual Voltage 13 Polarity Pulse Selection 14 Voltage Set Point Adjustment 14 Charge and Trigger a Pulse 14 Pulse Verification Procedure 15 Energy Meas...

Page 5: ...iii Cleaning 26 Calibration Information 26 Technical Assistance 27...

Page 6: ...iv...

Page 7: ...ical fundamentals clearly and be aware that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow...

Page 8: ...test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of the Tester and the equipment under test the operator should turn the rear panel pow...

Page 9: ...re when using the MegaPulse unit Pressing the POLARITY switch on the front panel can change the polarity of the output waveform The polarity is Normal when the NOR indicator is lit In this case the hi...

Page 10: ...tage Test Lead Red HVLR Qty 3 High Voltage Test Lead Black HVLB Qty 2 High Voltage Jumper Red HVLEC13 Option 100X includes one Jumper and option 34 also includes one Jumper Total Qty 2 if all options...

Page 11: ...e certified or approved by any recognized national test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument...

Page 12: ...Set point adjusted by frontal VOLTAGE knob or by PC TestMinder Polarity Control Positive and Negative Alternating control available with optional computer control Voltage Display 4 Digit LED Display V...

Page 13: ...th the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 1 and described in Table 1 The rear panel features of the MegaPulse are shown in Figu...

Page 14: ...rs will be blinking if the Interlock Switch is open 6 TRIGGER switch Triggers the output impulse waveform The impulse waveform will appear across the output leads 7 TRIGGER indicator This red indicato...

Page 15: ...unded 19 AUX OUTPUT indicator This Green indicator is lit to show that AUX OUPUT switch is set on the FLOAT position 20 S3 switch Used for common mode testing of figure 201 103 S3 switch should be clo...

Page 16: ...10 Figure 2 Controls Indicators Connectors MegaPulse D5 PF Rear Panel...

Page 17: ...ir flow to cool down the tester CAUTION When installing equipment make sure not to cover the fans and have it on a proper ventilated area 4 Appliance Inlet Fuse holder Power Switch Use supplied cord s...

Page 18: ...12 General High Voltage Electrical Diagram Figure 3 General High Voltage Electrical Diagram MegaPulse D5 PF...

Page 19: ...has disabled the keyboard or Voltage Knob it is possible to enable them by using the next keyboard sequence 1 Turn OFF the MegaPulse P tester 2 Press and hold the TRIGGER and NOR REV buttons 3 Turn ON...

Page 20: ...ng for a few seconds 2 While the digits are still blinking turn the VOLTAGE Adjust knob clockwise or counterclockwise to change the value of the blinking digits 3 While digits are still blinking press...

Page 21: ...mper on the front panel 6 Disconnect all high voltage cables from the outputs 50 ohm 400 ohm Energy Measurement Point and 50k ohm if included 7 Turn the rear panel Power Switch ON If residual voltage...

Page 22: ...RIGGER button to deliver the pulse be sure not to touch the output and return leads when pressing the trigger switch 20 Verify the amplitude of the negative pulse captured on the digital oscilloscope...

Page 23: ...17 Figure 7 Y1 Y2 Test Output...

Page 24: ...or value R of 100 1 which will result in a value between 99 and 101 We recommend using a value of 100 during energy calculations as a good approximation of the value for R over the test period Energy...

Page 25: ...e used to verify that the Applied Part or Chassis of the Equipment Under Test does not exceed 1 Volt peak in accordance with IEC EN UL 60601 1 A marginally failing product will cause the voltage on th...

Page 26: ...t the time at 500ms e Activate the Math function and turn off channel 1 and 2 9 Verify or adjust the Voltage Set Point to 5000V 10 Press the CHARGE button to start charging the internal high voltage c...

Page 27: ...ting parameters of IEC60601 They can be used as a key for the inputs and outputs of the MegaPulse D5 PF Note the actual connections on the front panel of the D5 PF have been added to Figure 50 of IEC...

Page 28: ...22 NOTE All D5 PF meet the specifications of IEC 60601 2 27 ed 3 0 where a 390K resistor replaced the 470K Figure 12 Example of D5 PF Connections on Figures 9A and 9B of IEC60601...

Page 29: ...put impedance of 50kohm and switch to select trigger duration of 200ms or 20 seconds this resistance values are need when testing to IEC 60601 2 34 Ed 2 0 If testing IEC 60601 2 34 Ed 3 0 no need of a...

Page 30: ...the Energy Measurement and Ground connectors 13 Turn ON the MegaPulse D5 PF 14 Tester is ready to CHARGE and TRIGGER with 100X Option To return the D5 PF tester to normal operation condition follow t...

Page 31: ...ger the pulse per the directions in the user manual 6 After the pulse put the S2 switch in the SINE WAVE position to apply the sine wave Vacuum Trigger Relay WARNING For this option never exceed 20 se...

Page 32: ...er General Maintenance To avoid contaminating the PWB with oil from your fingers handle it by the edges or wear gloves If the PWB becomes contaminated refer to the cleaning procedures given later in t...

Page 33: ...st USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA 921...

Reviews: