The  D5-PF  Medical  Defibrillation  tester  consists  of  two  parts;  the  pulse  generation  portion  and  the  measurement 
portion.  Please refer to the general electric circuit shown in Figure 3 to show the function of the D5-PF.    
 
The following procedure will explain how to generate a high voltage pulse with the MegaPulse unit.  

 

External Interlock 

 

The external interlock is a two-position terminal block located on the rear panel. When the External Interlock is open: 
 

 

TRIGGER and CHARGE lights will be blinking. 

 

The front keyboard will be disabled.  

 

If  the  unit  is  connected  to  a  computer  with  the  TestMinder  software,  the  interlock  status  will  be 
shown  on  the  computer,  condition  that  will  also  disable  polarity  changes,  charge,  and  trigger 
conditions.  

 

 

When the Interlock is closed, it enables all normal operations of the MegaPulse D5-PF features. 

 

 

Front Keyboard and Voltage Knob Enable 

 
If the MegaPulse D5-PF tester has disabled the keyboard or Voltage Knob, it is possible to enable them by using the 
next keyboard sequence: 
 

1.

 

Turn OFF the MegaPulse P tester. 

2.

 

Press and hold the

 TRIGGER 

and

 NOR-REV 

buttons

.

 

3.

 

Turn ON the MegaPulse tester. 

4.

 

Wait until the display shows rESE. 

5.

 

Release the TRIGGER and NOR-REV buttons. 

 

Discharging Residual Voltage 

 

Note that the Voltage meter may indicate that some residual voltage is present on the main storage capacitor, even 
when the MegaPulse tester is first turned ON.  This is due to inherent charging of the internal capacitors.  The following 
procedure will explain how to discharge a remaining voltage stored on the main internal capacitor.  

 

CAUTION 

High  voltage  generated  by  the  MegaPulse  tester  is  exposed  during  this  test.  A  risk  of  shock  exists.  
Exercise care when using the MegaPulse unit. 

 

1.

 

Turn the rear-panel Power Switch OFF. 

2.

 

Set AUX OUTPUT switch to “Return” position. 

3.

 

Set S3 switch to “Open” position. 

4.

 

If the MegaPulse includes option 100X, install the 100X jumper on the rear panel. 

5.

 

If the MegaPulse includes option 34, set the Trigger time switch to 200mS, install the 50k ohm jumper on 
the front panel. 

Summary of Contents for MegaPulse D5-PF

Page 1: ......

Page 2: ......

Page 3: ...nd ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this man...

Page 4: ...eyboard and Voltage Knob Enable 13 Discharging Residual Voltage 13 Polarity Pulse Selection 14 Voltage Set Point Adjustment 14 Charge and Trigger a Pulse 14 Pulse Verification Procedure 15 Energy Meas...

Page 5: ...iii Cleaning 26 Calibration Information 26 Technical Assistance 27...

Page 6: ...iv...

Page 7: ...ical fundamentals clearly and be aware that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow...

Page 8: ...test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of the Tester and the equipment under test the operator should turn the rear panel pow...

Page 9: ...re when using the MegaPulse unit Pressing the POLARITY switch on the front panel can change the polarity of the output waveform The polarity is Normal when the NOR indicator is lit In this case the hi...

Page 10: ...tage Test Lead Red HVLR Qty 3 High Voltage Test Lead Black HVLB Qty 2 High Voltage Jumper Red HVLEC13 Option 100X includes one Jumper and option 34 also includes one Jumper Total Qty 2 if all options...

Page 11: ...e certified or approved by any recognized national test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument...

Page 12: ...Set point adjusted by frontal VOLTAGE knob or by PC TestMinder Polarity Control Positive and Negative Alternating control available with optional computer control Voltage Display 4 Digit LED Display V...

Page 13: ...th the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 1 and described in Table 1 The rear panel features of the MegaPulse are shown in Figu...

Page 14: ...rs will be blinking if the Interlock Switch is open 6 TRIGGER switch Triggers the output impulse waveform The impulse waveform will appear across the output leads 7 TRIGGER indicator This red indicato...

Page 15: ...unded 19 AUX OUTPUT indicator This Green indicator is lit to show that AUX OUPUT switch is set on the FLOAT position 20 S3 switch Used for common mode testing of figure 201 103 S3 switch should be clo...

Page 16: ...10 Figure 2 Controls Indicators Connectors MegaPulse D5 PF Rear Panel...

Page 17: ...ir flow to cool down the tester CAUTION When installing equipment make sure not to cover the fans and have it on a proper ventilated area 4 Appliance Inlet Fuse holder Power Switch Use supplied cord s...

Page 18: ...12 General High Voltage Electrical Diagram Figure 3 General High Voltage Electrical Diagram MegaPulse D5 PF...

Page 19: ...has disabled the keyboard or Voltage Knob it is possible to enable them by using the next keyboard sequence 1 Turn OFF the MegaPulse P tester 2 Press and hold the TRIGGER and NOR REV buttons 3 Turn ON...

Page 20: ...ng for a few seconds 2 While the digits are still blinking turn the VOLTAGE Adjust knob clockwise or counterclockwise to change the value of the blinking digits 3 While digits are still blinking press...

Page 21: ...mper on the front panel 6 Disconnect all high voltage cables from the outputs 50 ohm 400 ohm Energy Measurement Point and 50k ohm if included 7 Turn the rear panel Power Switch ON If residual voltage...

Page 22: ...RIGGER button to deliver the pulse be sure not to touch the output and return leads when pressing the trigger switch 20 Verify the amplitude of the negative pulse captured on the digital oscilloscope...

Page 23: ...17 Figure 7 Y1 Y2 Test Output...

Page 24: ...or value R of 100 1 which will result in a value between 99 and 101 We recommend using a value of 100 during energy calculations as a good approximation of the value for R over the test period Energy...

Page 25: ...e used to verify that the Applied Part or Chassis of the Equipment Under Test does not exceed 1 Volt peak in accordance with IEC EN UL 60601 1 A marginally failing product will cause the voltage on th...

Page 26: ...t the time at 500ms e Activate the Math function and turn off channel 1 and 2 9 Verify or adjust the Voltage Set Point to 5000V 10 Press the CHARGE button to start charging the internal high voltage c...

Page 27: ...ting parameters of IEC60601 They can be used as a key for the inputs and outputs of the MegaPulse D5 PF Note the actual connections on the front panel of the D5 PF have been added to Figure 50 of IEC...

Page 28: ...22 NOTE All D5 PF meet the specifications of IEC 60601 2 27 ed 3 0 where a 390K resistor replaced the 470K Figure 12 Example of D5 PF Connections on Figures 9A and 9B of IEC60601...

Page 29: ...put impedance of 50kohm and switch to select trigger duration of 200ms or 20 seconds this resistance values are need when testing to IEC 60601 2 34 Ed 2 0 If testing IEC 60601 2 34 Ed 3 0 no need of a...

Page 30: ...the Energy Measurement and Ground connectors 13 Turn ON the MegaPulse D5 PF 14 Tester is ready to CHARGE and TRIGGER with 100X Option To return the D5 PF tester to normal operation condition follow t...

Page 31: ...ger the pulse per the directions in the user manual 6 After the pulse put the S2 switch in the SINE WAVE position to apply the sine wave Vacuum Trigger Relay WARNING For this option never exceed 20 se...

Page 32: ...er General Maintenance To avoid contaminating the PWB with oil from your fingers handle it by the edges or wear gloves If the PWB becomes contaminated refer to the cleaning procedures given later in t...

Page 33: ...st USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA 921...

Reviews: