11
ITEM NO.
NAME
FUNCTION
1
Ground for Energy Measurement
This jack is used along with the ENERGY MEASUREMENT jack to calculate the energy. This jack is
referenced to the chassis of the MegaPulse, and is referenced to earth ground as long as the MegaPulse is
properly grounded. Even though this jack is referenced to ground, it should be treated as HAZARDOUS
whenever the MegaPulse is turned ON.
2
Energy Measurement jack
This jack is necessary to verify the value of the 100 ohm resistor. The 100 ohm resistor value is measured
between the ENERGY MEASUREMENT jack and the GROUND jack on the rear panel.
3
Fans
The Fans on the rear panel maintain the air flow to cool down the tester. CAUTION: When installing
equipment, make sure not to cover the fans and have it on a proper ventilated area.
4
Appliance Inlet / Fuse holder /
Power Switch
Use supplied cord set to connect the MegaPulse D5-PF tester to an appropriate source of supply. Fuse
holder provides access for Fuse replacement, and the Power Switch is used to turn the tester ON and OFF.
5
TestMinder
Allows the communication between the tester and computer interface; a RS-232 to USB.
6
Interlock Switch
Emergency Stop
Close: Enables the tester buttons for operation.
Open: Stops any process in the tester and disables the buttons. The
TRIGGER
and
CHARGE
Indicators
will be blinking
7
Fuse replacement warning /
Rating of power supply
Specifies replacement fuse and required supply voltage.
Table 2. Control, Indicators, Connectors – MegaPulse D5-PF Rear Panel