Table 2. Controls, Indicators, Connectors – MegaPulse series Front Panel
ITEM
NAME
FUNCTION
1
VOLTAGE
Adjust Knob
Turn Clockwise to increase the output voltage after the
CHARGE
button has been pressed. Turn the knob
fully counterclockwise (lowest voltage setting) before the start of each test, and after the end of each test.
2
POLARITY
switch
The polarity switch only operates when the
CHARGE
indicator is lit, i.e. the output is not charged. The
polarity of the output waveform can be changed by pressing the
POLARITY
switch on the front panel. Press
this switch to toggle the output polarity from Normal to Reverse. The polarity is Normal when the
NOR
indicator is lit. In this case, the High Voltage will appear on the
OUTPUT
as a positive pulse relative to the
RETURN
jack. When the polarity switch is in the Reverse position (
REV
indicator is lit), the High voltage
will appear on the
OUTPUT
as a negative pulse relative to the
RETURN
jack.
3
NOR REV
indicator
Indicates the state of the Output Polarity switch.
NOR
indicate Normal position,
REV
indicates Reverse
position.
4
CHARGE
switch
Press this switch to begin charging the impulse storage capacitor (C1 in Figure 1). The
CHARGE
indicator
will turn off after the
CHARGE
switch is pressed, and the
TRIGGER
indicator will turn on. The voltage on
the capacitor will appear on the LED Display, Item 8. This voltage will appear across the output leads when
the
TRIGGER
switch is pressed. Note that the
POLARITY
switch is prevented from operating after the
CHARGE
switch has been pressed.
5
CHARGE
indicator
This Yellow indicator is lit to show that pressing the
CHARGE
switch is the next logical step in a test
sequence. Pressing the Charge switch causes the
CHARGE
indicator to go out.
6
TRIGGER
switch
Press this switch (after pressing the
CHARGE
switch to charge the storage capacitor) to trigger the output
impulse waveform. The impulse waveform will appear across the output leads.
7
TRIGGER
indicator
This Red indicator is lit to show that pressing the
TRIGGER
switch is the next logical step in a test sequence.
This indicator will turn on after the
CHARGE
switch is pressed, and will remain on until the
TRIGGER
switch is pressed. Pressing the
TRIGGER
switch causes the
TRIGGER
indicator to go out.
8
Voltage meter
Displays the output voltage set point. This voltage is the open-circuit peak voltage that will appear across the
output leads when the trigger button is pressed. The voltage reading will increase from zero to the voltage set
point when the Charge switch is pressed. Note that the Voltage meter may indicate that some residual voltage
is present on the main storage capacitor, even when the tester is first turned ON. This is due to inherent
charging of the internal capacitors. Pressing the
TRIGGER
switch will discharge the capacitors (be sure not
to touch the output and return leads when pressing the trigger switch). Note that the peak amplitude of the
measured output waveform is proportional to the voltage that is read of the front panel of the MegaPulse, but it
will always be somewhat lower. This is because the meter on the MegaPulse is measuring the voltage on the
main impulse storage capacitor (C1 in Figure 1). This voltage will intentionally dissipate to some extent before
reaching the output leads. Therefore, it is important to measure the peak amplitude of the output waveform,
and adjust the output of the MegaPulse accordingly.
9
OUTPUT
jack
The impulse waveform appears on the
OUTPUT
jack, referenced to the
RETURN
jack. When the
POLARITY
switch is in the Normal position (
NOR
indicator is lit) the output will be a positive pulse. When
the
POLARITY
switch is in the Reverse position (
REV
indicator is lit) the output will be a negative pulse.
10
RETURN
jack
This is the return for the impulse waveform. This jack is referenced to the chassis of the MegaPulse, and is
referenced to earth ground as long as the MegaPulse is properly grounded. Even though this jack is referenced
to ground, it should be treated as hazardous whenever the MegaPulse is turned ON.
11
Model designation area
The specific MegaPulse model is indicated in this area. Optional metering/measurement test points may also
be installed in this area.
Summary of Contents for MegaPulse 2x10s-2.5
Page 1: ...i MegaPulse series Impulse Tester Instruction Manual COMPLIANCE WESTUSA ...
Page 4: ...iv ...
Page 8: ......
Page 12: ...8 Figure 2 Controls Indicators Connectors MegaPulse series Front Panel ...
Page 14: ...10 Figure 3 Controls Indicators Connectors MegaPulse series Rear Panel ...
Page 16: ...12 Figure 4 Controls Indicators Connectors MegaPulse series Breakdown Detection ...
Page 22: ...18 Figure D 1 Defib 5 Connections 1 ...
Page 23: ...19 Figure D 2 Y1 Y2 Test Output 1 ...
Page 24: ......
Page 26: ......
Page 32: ...28 ...