15
shunt.
12.
Connect the measuring instrument
across the current shunt, keeping in mind
that the Return lead is referenced to
ground.
13.
Turn the Tester on. Toggle the
POLARITY
switch if necessary so that
the
NOR
indicator is lit.
14.
Push the yellow
CHARGE
button.
Verify the red
TRIGGER
indicator is
now lit.
15. Adjust
the
VOLTAGE
knob so that the
front panel LED display is reading a
voltage that is suitable for the measuring
instrument that is being used. Push the red
TRIGGER
button, and view the resulting
impulse waveform on the measuring
instrument.
16.
Repeat steps 12 through 14, except this
time toggle the
POLARITY
switch so
that the
REV
indicator is lit. Note that the
impulse waveform will now be a negative
pulse, so it will probably be necessary to
make adjustments to the measuring
instrument to get a proper reading.
16. Adjust
the
VOLTAGE
knob fully
counterclockwise, to the minimum
position. Turn the rear-panel power
switch OFF.
Testing
This section describes how the MegaPulse Tester
is used to conduct a test. The
test
can
be
stopped immediately at any time by turning OFF
the rear-panel power switch.
1.
Connect the Tester to a proper source of
supply using the included 18 AWG power
supply cord. Make sure that the front
panel
VOLTAGE
adjust know is turned
fully counterclockwise.
2.
Plug the Output and Return test leads in
to the jacks on the front panel.
3.
Connect the ends of the test leads to the
equipment under test.
4.
Connect the ends of the test leads to the
optional 1000:1 voltage sense jack on the
front panel (if provided). Alternatively,
connect an appropriate measuring
instrument (typically an oscilloscope with
a high-voltage probe) to the equipment
under test in order to monitor the results of
the impulse test. Note that the
RETURN
lead is referenced to the chassis of the
MegaPulse tester.
5.
Turn the Tester on. Toggle the
POLARITY
switch as needed so that the
NOR
or
REV
indicator is lit.
6.
Note that the Voltage meter may
indicate that some residual voltage is
present on the main storage capacitor,
even when the tester is first turned ON.
This is due to inherent charging of the
internal capacitors. Pressing the
TRIGGER
switch will discharge the
capacitors (be sure not to touch the output
and return leads when pressing the trigger
switch).
7.
Push the yellow
CHARGE
button.
Verify the red
TRIGGER
indicator is
now lit.
8. Adjust
the
VOLTAGE
knob so that the
front panel LED display is reading a
voltage that is suitable for the measuring
instrument that is being used. Push the red
TRIGGER
button, and view the resulting
impulse waveform on the measuring
instrument.
9.
Note that the peak amplitude of the
measured output waveform is
proportional to the voltage that is read of
the front panel of the MegaPulse, but it
will always be somewhat lower. This is
because the meter on the MegaPulse is
measuring the voltage on the main
impulse storage capacitor (C1 in Figure
1). This voltage will intentionally
dissipate to some extent before reaching
the output leads. Therefore, it is
important to measure the peak amplitude
of the output waveform, and adjust the
output of the MegaPulse accordingly.
Summary of Contents for MegaPulse 2x10s-2.5
Page 1: ...i MegaPulse series Impulse Tester Instruction Manual COMPLIANCE WESTUSA ...
Page 4: ...iv ...
Page 8: ......
Page 12: ...8 Figure 2 Controls Indicators Connectors MegaPulse series Front Panel ...
Page 14: ...10 Figure 3 Controls Indicators Connectors MegaPulse series Rear Panel ...
Page 16: ...12 Figure 4 Controls Indicators Connectors MegaPulse series Breakdown Detection ...
Page 22: ...18 Figure D 1 Defib 5 Connections 1 ...
Page 23: ...19 Figure D 2 Y1 Y2 Test Output 1 ...
Page 24: ......
Page 26: ......
Page 32: ...28 ...