2
Safety Techniques
The high voltage circuit of the MegaPulse can be shut off at any time by turning OFF the rear
power switch. Note that there are residual voltages inside the equipment that may take up to 5 minutes to
bleed off to safe levels.
The MegaPulse tester is provided with a
VOLTAGE ADJUST
knob on the front panel. This
should always be adjusted fully counterclockwise to the minimum position at the start of testing. In
addition, this knob should be adjusted back to the minimum position at the completion of all testing.
The MegaPulse tester is provided with a
CHARGE
switch that is in the unarmed "Standby"
setting when the tester is first turned ON. When the yellow
CHARGE
button is lit, the tester will not
provide high voltage until the
CHARGE
Button and the
TRIGGER
Button have been pressed in order.
To prevent inadvertent operation, the operator should be instructed not to press the
CHARGE
Button until
the test is ready.
The MegaPulse tester has been designed for one-touch operation with the right hand. If possible,
it should be set up to the left and in front of the equipment under test. The equipment under test should be
connected to the MegaPulse tester and then left alone by the operator. After the operator is clear of the
Tester and the equipment under test, the operator should turn the rear-panel power switch to ON, press the
CHARGE
Button, adjust the voltage to the desired level (as displayed on the front panel meter), then press
the
TRIGGER
Button, with his right hand. This will allow the greatest separation between the operator
and the test being conducted.
The MegaPulse tester is designed to bleed the high voltage away after the test has concluded. In
order to ensure that any voltage present in the equipment being tested has been completely bled away, the
operator should not unplug the equipment under test from the MegaPulse until the front panel meter reads a
safe level (40 volts or less is generally considered a safe level). Pressing the TRIGGER button before
disconnecting main power (or turning the equipment off) will ensure that the internal capacitors are
discharged as much as possible.
Using the MegaPulse Impulse Tester
The impulse withstand test involves high voltage and caution should be exercised when using the
Tester. The
RETURN
lead is referenced to building ground when properly connected. However, both the
OUTPUT
and
RETURN
leads must always be treated as Hazardous whenever the power switch of the
MegaPulse is in the ON position.
The MegaPulse impulse tester generates the impulse waveform only; it does not determine
Passing or Failing results. It is Operator’s responsibility to monitor the output waveform and determine
Passing or Failing results. In monitoring the impulse waveform, consider the following points:
The Impulse waveform is high voltage, and high frequency (short duration). Always ensure that
the measuring instrument (usually an oscilloscope with a high-voltage probe) is rated for the voltage
involved, and that the frequency response of the instrument and probe are capable of measuring the output
waveform of the MegaPulse Impulse Tester. A measuring instrument or probe with a low frequency
response will result in erroneous readings that could be mis-read.
Pressing the POLARITY switch on the front panel can change the polarity of the output
waveform. The polarity is Normal when the
NOR
indicator is lit. In this case, the High Voltage will
appear on the
OUTPUT
as a positive pulse relative to the
RETURN
jack. When the polarity switch is in
the Reverse position (
REV
indicator is lit), the High voltage will appear on the
OUTPUT
as a negative
pulse relative to the
RETURN
jack. The polarity switch only operates when the
CHARGE
LED is lit, i.e.
the output is not charged.
Note that the Voltage meter may indicate that some residual voltage is present on the main
storage capacitor, even when the tester is first turned ON. This is due to inherent charging of the internal
capacitors. Pressing the
TRIGGER
switch will discharge the capacitors (be sure not to touch the output
and return leads when pressing the trigger switch).
Summary of Contents for MegaPulse 2x10s-2.5
Page 1: ...i MegaPulse series Impulse Tester Instruction Manual COMPLIANCE WESTUSA ...
Page 4: ...iv ...
Page 8: ......
Page 12: ...8 Figure 2 Controls Indicators Connectors MegaPulse series Front Panel ...
Page 14: ...10 Figure 3 Controls Indicators Connectors MegaPulse series Rear Panel ...
Page 16: ...12 Figure 4 Controls Indicators Connectors MegaPulse series Breakdown Detection ...
Page 22: ...18 Figure D 1 Defib 5 Connections 1 ...
Page 23: ...19 Figure D 2 Y1 Y2 Test Output 1 ...
Page 24: ......
Page 26: ......
Page 32: ...28 ...