7
Section 4
Troubleshooting the HTT-1
The HTT-1 does not require calibration. If the HTT-1 does not operate as described in the
Theory of Operation section, and the equipment connected to the HTT-1 has been verified to be
in good working condition, then the HTT-1 should be returned to Compliance West USA for
inspection and/or repair.
Table 3 shows expected results for a known good hipot tester for all recommended operating
modes.
Switch designation / Switch position
GROUND
LEAKAGE
HIPOT
Operating mode
PASS
PASS
PASS
No failure: the tester connected to the HTT-1
should indicate test completion with a passing
result.
FAIL
PASS
PASS
Ground Fail: the tester connected to the HTT-1
should indicate a “ground open” or “ground
fail” condition.
PASS
FAIL
PASS
Leakage Fail: the tester connected to the HTT-
1 should indicate an “excess leakage” or
“excess current” failure.
PASS
PASS
FAIL
Hipot Fail: The tester connected to the HTT-1
should indicate a “hipot fail” or “dielectric
failure”. The high-voltage arc that occurs to
create the simulated test failure is visible
through the ARC VIEW window.
Table 3: HTT-1 Normal operating modes and expected results.