background image

 

 

Section 4 

Troubleshooting the HTT-1 

 

The HTT-1 does not require calibration.  If the HTT-1 does not operate as described in the 
Theory of Operation section, and the equipment connected to the HTT-1 has been verified to be 
in good working condition, then the HTT-1 should be returned to Compliance West USA for 
inspection and/or repair. 
 
Table 3 shows expected results for a known good hipot tester for all recommended operating 
modes. 

 
 

Switch designation / Switch position 

GROUND 

LEAKAGE 

HIPOT 

Operating mode 

PASS 

PASS 

PASS 

No failure: the tester connected to the HTT-1 

should indicate test completion with a passing 

result. 

FAIL 

PASS 

PASS 

Ground Fail: the tester connected to the HTT-1 

should indicate a “ground open” or “ground 

fail” condition. 

PASS 

FAIL 

PASS 

Leakage Fail:  the tester connected to the HTT-

1 should indicate an “excess leakage” or 

“excess current” failure. 

PASS 

PASS 

FAIL 

Hipot Fail:  The tester connected to the HTT-1 

should indicate a “hipot fail” or “dielectric 

failure”.  The high-voltage arc that occurs to 

create the simulated test failure is visible 

through the ARC VIEW window. 

 

Table 3:  HTT-1 Normal operating modes and expected results. 

Summary of Contents for HTT-1

Page 1: ...HTT 1 Function Tester Instruction Manual COMPLIANCE WEST USA...

Page 2: ...i...

Page 3: ...ons Compliance West USA is proud to present you with your HTT 1 The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and conf...

Page 4: ...ii...

Page 5: ...troduction and Specifications 2 Introduction 2 Specifications 2 Operation 3 Setting up the HTT 1 3 Front Panel Features 3 Setup Procedure Hipot Tester 3 Setup Procedure Ground Continuity Tester 4 Unan...

Page 6: ...a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1 to the tester outputs 2 Set all three switches on the HTT 1...

Page 7: ...fication tool The HTT 1 is warranted for a period of one year upon shipment to the original purchaser Specifications Specifications for the HTT 1 are listed in Table 1 below Table 1 Specifications of...

Page 8: ...continuity tester Because of the varying types of tester outputs cables are not provided with the HTT 1 For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1 a O...

Page 9: ...tester identify the return lead of the high current output This is connected to the black banana jack receptacle of the HTT 1 b On your ground continuity tester plug the HTT 1 into the high current re...

Page 10: ...5 Figure 1 HTT 1 Front Panel Features 6 5 4 3 1 2...

Page 11: ...breakdown occurred 4 Hipot Test Switch When set to FAIL a dielectric breakdown will be simulated and a spark can be seen in the Arc View window When set to PASS a normal passing result is simulated Us...

Page 12: ...ROUND LEAKAGE HIPOT Operating mode PASS PASS PASS No failure the tester connected to the HTT 1 should indicate test completion with a passing result FAIL PASS PASS Ground Fail the tester connected to...

Page 13: ...748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Monday through Friday Compliance West USA 2120 Jimmy Durante Blvd Suite 124 De...

Reviews: