Section 1 

Operating Instructions 

 

Theory of Operation 

 
 

The HTT-1 is designed to be a solution for fast, easy function testing of all dielectric 

withstand and continuity testers, including high current models delivering up to 30 amps.  It 
allows function tests of all three failure modes to ensure that your tester is functioning properly.   
 
 

The HTT-1 is not designed to verify calibration points except as noted in the 

Specifications Section of this manual.  Optional calibration point verification is available as an 
option. Contact the factory for details. 
 
 

The HTT-1, used daily, will prove function of the safety testers before shifts or before 

testing of a lot of products.  It is provided with three switches which simulate passing and failing 
equipment for ground continuity, high leakage between primary circuits and ground, and 
dielectric withstand failure between primary and ground circuits.  While all three switches can be 
simultaneously set for failing modes, we recommend that the HTT-1 be set with one failure at a 
time to ensure that your testers are functioning correctly.   
 
 

A proper test plan would be constituted of the following tests: 
1. 

Connect HTT-1 to the tester outputs. 

2. 

Set all three switches on the HTT-1 to ‘PASS’. 

3. 

Verify passing result on the Tester. 

4. 

Set the Ground switch on the HTT-1 to ‘FAIL’; set both other switches to 
‘PASS’. 

5. 

Verify Ground failure on the Tester. 

6. 

Set the Leakage switch on the HTT-1 to ‘FAIL’; set both other switches to 
‘PASS’. 

7. 

Verify Leakage current failure on the Tester. 

8. 

Set the Hipot switch on the HTT-1 to ‘FAIL’; set both other switches to ‘PASS’. 

9. 

Confirm high voltage output by watching for spark in Arc Detect window of 
HTT-1; verify dielectric breakdown failure on the Tester. 

 

The above plan is presented in more detail in the Operation Section of this Manual; 

showing proper connection of the HTT-1 to any dielectric tester and any high current ground 
continuity tester you may wish to test. 

 
If any of the tests above return unanticipated results, you may have a problem with your 

tester.  It is recommended that it be removed from service until the problem can be resolved.  
Section 4 of this manual also has troubleshooting information to help you determine if the HTT-1 
is functioning correctly. 

 

Summary of Contents for HTT-1

Page 1: ...HTT 1 Function Tester Instruction Manual COMPLIANCE WEST USA...

Page 2: ...i...

Page 3: ...ons Compliance West USA is proud to present you with your HTT 1 The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and conf...

Page 4: ...ii...

Page 5: ...troduction and Specifications 2 Introduction 2 Specifications 2 Operation 3 Setting up the HTT 1 3 Front Panel Features 3 Setup Procedure Hipot Tester 3 Setup Procedure Ground Continuity Tester 4 Unan...

Page 6: ...a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1 to the tester outputs 2 Set all three switches on the HTT 1...

Page 7: ...fication tool The HTT 1 is warranted for a period of one year upon shipment to the original purchaser Specifications Specifications for the HTT 1 are listed in Table 1 below Table 1 Specifications of...

Page 8: ...continuity tester Because of the varying types of tester outputs cables are not provided with the HTT 1 For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1 a O...

Page 9: ...tester identify the return lead of the high current output This is connected to the black banana jack receptacle of the HTT 1 b On your ground continuity tester plug the HTT 1 into the high current re...

Page 10: ...5 Figure 1 HTT 1 Front Panel Features 6 5 4 3 1 2...

Page 11: ...breakdown occurred 4 Hipot Test Switch When set to FAIL a dielectric breakdown will be simulated and a spark can be seen in the Arc View window When set to PASS a normal passing result is simulated Us...

Page 12: ...ROUND LEAKAGE HIPOT Operating mode PASS PASS PASS No failure the tester connected to the HTT 1 should indicate test completion with a passing result FAIL PASS PASS Ground Fail the tester connected to...

Page 13: ...748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Monday through Friday Compliance West USA 2120 Jimmy Durante Blvd Suite 124 De...

Reviews: