a)  Your hipot tester should indicate ‘FAIL’, stop testing, and sound a buzzer or otherwise 

warn the operator of a problem. 

5)  Conduct  a  simulated  HIPOT  FAILURE  test  by  setting  the  HIPOT  switch  on  the  HTT-1  to 

‘FAIL’.    (All  other  switches  on  the  HTT-1  should  be  set  to  ‘PASS’.)    Begin  testing  by 
starting your hipot tester. 
a)  The  dielectric  breakdown  will  be  visible  as  an  arc  in  the  ARC  VIEW  window  of  the 

HTT-1. 

b)  Your  hipot  tester  should  indicate  ‘FAIL’,  stop  testing  and  sound  a  buzzer  or  otherwise 

warn the operator of a problem.  (If the current of the spark is above the leakage limit, 
the Excess Leakage failure may occur as well) 

 

Setup Procedure:  Ground Continuity Tester 

 

This procedure will describe connection of the HTT-1 to a ground continuity tester.  Because of 
the varying types of tester outputs, cables are not provided with the HTT-1.  For custom cables, 
please contact Compliance West USA. 
 
1)  Connect the ground continuity tester to the HTT-1.   

a)  On your ground continuity tester, identify the return lead of the high current output.  This 

is connected to the black banana jack receptacle of the HTT-1.  

b)  On your ground continuity tester, plug the HTT-1 into the high current receptacle. 

2)  Conduct a simulated PASSING test by setting all three switches of the HTT-1 to the PASS 

(green)  setting  and  initiating  a  ground  continuity  test  by  starting  your  ground  continuity 
tester. 
a)  Your ground continuity tester should indicate ‘PASS’. 

3)  Conduct a simulated GROUND OPEN test by setting the GROUND switch on the HTT-1 to 

‘FAIL’.    (All  other  switches  on  the  HTT-1  should  be  set  to  ‘PASS’.)    Begin  testing  by 
starting your ground continuity tester. 

a) 

Your ground continuity tester should indicate ‘FAIL’, stop testing, and sound a buzzer or 
otherwise warn the operator of a problem.

 

 

Unanticipated Results 

 

The HTT-1 test may cause your tester to give results not consistent with those anticipated by the 
switch setting of the HTT-1.  Until the problem is resolved, your tester should be removed from 
service. 
 
If there is a question regarding the proper operation of the HTT-1, troubleshooting information in 
Section 4 can be used to verify proper operation of the HTT-1. 

 
 

Summary of Contents for HTT-1

Page 1: ...HTT 1 Function Tester Instruction Manual COMPLIANCE WEST USA...

Page 2: ...i...

Page 3: ...ons Compliance West USA is proud to present you with your HTT 1 The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and conf...

Page 4: ...ii...

Page 5: ...troduction and Specifications 2 Introduction 2 Specifications 2 Operation 3 Setting up the HTT 1 3 Front Panel Features 3 Setup Procedure Hipot Tester 3 Setup Procedure Ground Continuity Tester 4 Unan...

Page 6: ...a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1 to the tester outputs 2 Set all three switches on the HTT 1...

Page 7: ...fication tool The HTT 1 is warranted for a period of one year upon shipment to the original purchaser Specifications Specifications for the HTT 1 are listed in Table 1 below Table 1 Specifications of...

Page 8: ...continuity tester Because of the varying types of tester outputs cables are not provided with the HTT 1 For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1 a O...

Page 9: ...tester identify the return lead of the high current output This is connected to the black banana jack receptacle of the HTT 1 b On your ground continuity tester plug the HTT 1 into the high current re...

Page 10: ...5 Figure 1 HTT 1 Front Panel Features 6 5 4 3 1 2...

Page 11: ...breakdown occurred 4 Hipot Test Switch When set to FAIL a dielectric breakdown will be simulated and a spark can be seen in the Arc View window When set to PASS a normal passing result is simulated Us...

Page 12: ...ROUND LEAKAGE HIPOT Operating mode PASS PASS PASS No failure the tester connected to the HTT 1 should indicate test completion with a passing result FAIL PASS PASS Ground Fail the tester connected to...

Page 13: ...748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Monday through Friday Compliance West USA 2120 Jimmy Durante Blvd Suite 124 De...

Reviews: