background image

 

Section 2 

 

Introduction and Specifications 

 

Introduction 

 

This  manual  contains  complete  operating,  maintenance  and  calibration  instructions  for  the 
Compliance West USA Model HTT-1 Function Tester. 
 
The instrument is a portable function tester used to verify proper operation of hipot and ground 
continuity  testers.    Unless  optionally  equipped  as  shown  in  the  Specification  Section  of  this 
manual, the HTT-1 is not intended as a calibration verification tool.   
 
The HTT-1 is warranted for a period of one year upon shipment to the original purchaser. 
 

Specifications 

 

Specifications for the HTT-1 are listed in Table 1 below. 
 

 

 

 
 
 
 
 
 
 
 
 
 

 
 
 

 

Table 1:  Specifications of the HTT-1 

 

 

 

 

 
 
 
 
 
 

 

Specifications of the HTT-1  
 

Maximum current capability of GROUND test circuit:  30 Amps 
Setting of PASS for Ground Circuit: 

 

 

0 ohm 

Setting of FAIL for Ground Circuit: 

 

 

open circuit 

 

Setting of PASS for Leakage Current:   

 

open circuit 

Setting of FAIL for Leakage Current: 

 

 

short-circuit 

 

Maximum voltage capability of HIPOT test circuit: 

3000 V 

Setting of PASS for Hipot test:   

 

 

open circuit 

Setting of FAIL for Hipot test:    

 

 

arc-over at 750 V+400V

 

 

Summary of Contents for HTT-1

Page 1: ...HTT 1 Function Tester Instruction Manual COMPLIANCE WEST USA...

Page 2: ...i...

Page 3: ...ons Compliance West USA is proud to present you with your HTT 1 The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and conf...

Page 4: ...ii...

Page 5: ...troduction and Specifications 2 Introduction 2 Specifications 2 Operation 3 Setting up the HTT 1 3 Front Panel Features 3 Setup Procedure Hipot Tester 3 Setup Procedure Ground Continuity Tester 4 Unan...

Page 6: ...a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1 to the tester outputs 2 Set all three switches on the HTT 1...

Page 7: ...fication tool The HTT 1 is warranted for a period of one year upon shipment to the original purchaser Specifications Specifications for the HTT 1 are listed in Table 1 below Table 1 Specifications of...

Page 8: ...continuity tester Because of the varying types of tester outputs cables are not provided with the HTT 1 For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1 a O...

Page 9: ...tester identify the return lead of the high current output This is connected to the black banana jack receptacle of the HTT 1 b On your ground continuity tester plug the HTT 1 into the high current re...

Page 10: ...5 Figure 1 HTT 1 Front Panel Features 6 5 4 3 1 2...

Page 11: ...breakdown occurred 4 Hipot Test Switch When set to FAIL a dielectric breakdown will be simulated and a spark can be seen in the Arc View window When set to PASS a normal passing result is simulated Us...

Page 12: ...ROUND LEAKAGE HIPOT Operating mode PASS PASS PASS No failure the tester connected to the HTT 1 should indicate test completion with a passing result FAIL PASS PASS Ground Fail the tester connected to...

Page 13: ...748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Monday through Friday Compliance West USA 2120 Jimmy Durante Blvd Suite 124 De...

Reviews: