6
ELECTRICAL
Output
0-14000 volts DC
Leakage Current
1-5 mA
Pass/Fail Criteria:
Leakage Current:
Pass/Fail point user adjustable.
Dielectric Breakdown:
Separate high frequency detection circuit for
breakdown spike detection
Test Time:
User adjustable 1->60 sec., defeatable
Voltage Ramp-up Time:
User adjustable 1-5 sec., defeatable
Voltage Ramp-down Time:
Factory set 8 sec. maximum
Meter Accuracy
+/- 3 % full scale
Duty cycle
100 %
Test adjustments
Rear Panel:
Ramp Time
Test Time
Leakage Limit
Voltage Adjust
Voltage Ramp ON/DEFEAT
Timer Duration ON/DEFEAT
Failure Shutdown ON/DEFEAT
ENVIRONMENTAL
Operating Temperature
15-40
°
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
Model HT-10KVDC: 114-127 volts, 50/60 Hz
Weight
22 lbs.
RESET and TEST Lamp Type
Replace with type 73 14V lamp.
SAFETY AGENCY TOPICS
Transformer Output
< 500VA
Visual Indication of Voltage Output
Provided by front panel meter, directly
connected to high voltage output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LEDs on front panel
Test can be automatically terminated on failure
Leakage Test
Provided; 120Kohm calibration resistor
provided on rear panel.
Table 2-1 HT-10KVDC Specifications