34
7.5
Belt
Speed
Calibration
Recommended
calibration
of
the
treadmill
belt
speed
and
acceleration
parameters
should
be
performed
every
3
months.
The
calibration
procedure
is
as
follows:
1)
Ensure
the
Treadmill
Controller
front
panel
toggle
is
set
to
“STOP”.
2)
Place
speed
sensor
on
treadmill
belt
and
securely
fasten
with
screw
nut
in
position
shown
(F
‐
53).
3)
Using
the
dial
on
the
Treadmill
Controller
front
panel,
adjust
the
speed
to
88.7
m/min.
4)
Next,
set
the
electric
stimulus
frequency
to
2
Hz
on
Treadmill
Controller
front
panel.
5)
Push
the
“ODOMETER
RESET”
button
to
initialize
calibration
procedure.
The
LCD
on
the
front
panel
will
read
“AutoCal
Speed”
as
shown
(F
‐
54).
6)
Push
the
“ODOMETER
RESET”
button
a
second
time
to
advance
to
belt
calibration
start.
The
LCD
on
the
front
panel
will
read
“Run
to
Start…”
as
shown
(F
‐
55).
7)
Set
toggle
to
“RUN”
to
begin
the
series
of
belt
speed
and
acceleration
tests.
8)
Upon
belt
calibration
completion,
the
most
recent
calibration
is
automatically
saved,
and
speed
sensor
counts
(OD
counts)
are
displayed.
Obtaining
count
values
between
4000
‐
5000
indicates
calibration
was
successful
(F
‐
56).
9)
To
exit
the
calibration,
press
the
“ODOMETER
RESET
“button.
A
“Calibration
Saved”
message
will
be
displayed
as
a
further
reassurance
that
the
most
recent
calibration
was
saved
automatically.
To
complete
the
process
and
easily
return
to
the
main
menu,
simply
power
the
unit
off
and
then
back
on
using
the
switch
located
on
the
rear
of
the
controller.
Figure
53
Figure
54
Figure
55
Figure
56
Summary of Contents for 0184-006M
Page 2: ......