![Colin BP-S510 Operation Manual Download Page 78](http://html1.mh-extra.com/html/colin/bp-s510/bp-s510_operation-manual_2637159078.webp)
72
Label
Name of each channel (P1 or P2) can be selected in this menu.
Pressure Zero Setting
1. Three way tap for zero calibration should be set at the same level as patient’s heart.
9
Make sure if there is no bubble in the each part.
9
Make sure if the drop volume of the IV set is at the proper position.
2. The handle of the three way tap should be to the patient so that the pressure to the
transducer is open to the atmosphere level.
9
Please make sure if release part to the atmosphere in not covered by cap.
3. Set the pressure zero setting to
Yes
. “Zero CAL in progress” message will appear.
4. When the message “Zero CAL in progress” is disappeared and measurement will be
initiated.
5. When Zero Calibration is done, the handle of three way tap should be turned to the
atmosphere direction and the release part should be covered by the cap.
6. IBP measurement can start by changing the lever of three way tap so as to transfer the
pressure from the catheter to transducer.
Note: If there is bubble in the any part, you can remove the bubble by controlling the
transducer.
Note: If zero calibration can not be done correctly, the readings will not be accurate.
Note: When zero calibration is done, make sure that release port of three way tap for
Zero calibration is open, not covered by the cap of injector.
Description of IBP Waveform Menu Functions
1 IBP 1 waveform icon
4
IBP 2 waveform icon
2 IBP 1 scale bar
5
IBP 2 scale bar
3 IBP 1 waveform
6
IBP 2 waveform
Figure 42. IBP Waveform Display
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...