![Colin BP-S510 Operation Manual Download Page 109](http://html1.mh-extra.com/html/colin/bp-s510/bp-s510_operation-manual_2637159109.webp)
103
EXTERNAL INTERFACE
General
The monitor provides external connectors on the right panel to support communication
with external equipment and functions such as a nurse call, software upgrades or PC
connection.
WARNING: Any connections between this monitor and other devices must
comply with applicable medical systems safety standards such as IEC 60601-1.
Failure to do so could result in unsafe leakage current and grounding conditions.
Note: This equipment is to be used on a network and the communication wirings (RJ45
LAN and RS232) are limited to inside of the building.
Cable Connections
RS-232
The pin layouts of 15-pin RS-232 interface are illustrated below.
1
12
11
10
9
5
4
3
2
6 7
8
13 14 15
Figure 58. Data Port Pin Layout
Table 40. RS-232 Serial Interface Connections
Pin #
Signal
1
Not used (open)
2 RS232
RX
3 RS232
TX
4
not used (open)
5 RS232
ground
6
not used (open)
7
Nurse call normally open
8
Nurse call normally closed
9
not used (open)
10 RS232
ground
11
not used (open)
12
not used (open)
13
not used (open)
14
not used (open)
15 Nurse
call
common
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...