60
Description of SpO2 Menu Functions
1 SpO
2
icon
4
SpO
2
alarm limits value
2 Pulse amplitude indicator
5
SpO
2
value
3 SpO
2
alarm icon
Figure 29. SpO
2
Display
Figure 30. SpO
2
Menu
Table 27. SpO
2
Menu
Level 1 Menu
Level 2 Menu or Response
SpO
2
MENU
C-Lock On,
Off
(Alarm Limits Adjustment)
▲
Upper Alarm Limit
▼
Lower Alarm Limit
SpO
2
Alarm Suspend
On, Off
Return
C-Lock
When C-Lock is turned on in the SpO
2
menu, C-Lock automatically becomes operational
any time a valid ECG signal is detected by the monitor. It is not necessary to turn C-Lock
off if an ECG signal is not available; the monitor handles this function automatically. If the
ECG signal is noisy, or of poor quality, SpO
2
performance may be improved by turning C-
Lock off. C-Lock provides ECG synchronization for more reliable saturation
measurements. An ECG (R-wave) signal can be used as a time reference to identify the
pulse and synchronize saturation measurements. C-Lock enhances performance while
maintaining rapid response time.
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...