Audible Alarm Silence
WARNING: Do not silence the audible alarm or decrease its volume if patient
safety could be compromised.
When an alarm occurs, you can silence the audible alarm for audible alarm silence period
(30, 60, 90 or 120 seconds) selected via service menu. However, visual alarms continue
during this time. The factory default of audible alarm silence period is 120 seconds.
To silence an audible alarm:
1. Press
Alarm silence button
to immediately silence the alarm tone. The alarm resumes
after the audible alarm silence period if the alarm condition has not been corrected.
2. Check the patient and provide appropriate care.
During the audible alarm silence period, you can press
Alarm silence button
again to re-
enable the audible alarm tones. Also, if another alarm occurs during the audible alarm
silence period, the audible alarm tones will be automatically re-enabled.
Figure 19.Audible Alarm Silence Display
Note: The audible alarms caused by some technical errors may be canceled by pressing
Alarm silence button
. However, battery failure and physiological alarms cannot be
canceled until the alarm condition is corrected.
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...