FACTORY DEFAULTS
General
The monitor is shipped with factory default settings. Authorized personnel can use the
procedures described in the service manual to change default settings.
Parameter Ranges and Default Settings
Table 41. Parameter Ranges and Factory Defaults
Factory Defaults
Parameter
Ranges/Selections
Adult
Neonatal
ECG
ECG Cable Select
3 Leads, 5 Leads, AUTO
AUTO
AUTO
ECG Lead Select
I, II, III, aV
R
, aV
L
, aV
F
, V(Chest Lead)
-
-
ECG Size (mm/mV)
×1/4, ×1/2, ×1, ×1.5, ×2
×1
×1
ECG Filter Mode
Monitor, Low Extend, Filter,
Respiration Rejection
Monitor Monitor
ECG Pacer Detect
On, Off
Off
Off
ECG Sweep Speed
12.5, 25.0, 50.0 mm/s
25.0 mm/s
25.0 mm/s
HR/PR Source
AUTO, HR, PR
AUTO
AUTO
HR/PR Upper Alarm Limits
35 to 305 BPM (Adult/Neo) (5 BPM steps)
180 BPM
200 BPM
HR/PR Lower Alarm Limits
30 to 300 BPM (Adult/Neo) (5 BPM steps)
40 BPM
50 BPM
NIBP
NIBP Initial Cuff Inflation
120, 140, 160, 180, 200, 220mmHg (Adult)
(16.0, 18.7, 21.3, 24.0, 26.7, 29.3, kPa)
80, 100, 120, 140 mmHg (Neo)
(9.3, 12.0, 14.7, 16.0, 18.7 kPa)
180 mmHg
24.0 kPa
120 mmHg
16.0 kPa
BP On Alarm
On, Off
Off
Off
Smart Clock
On, Off
On
On
Smart Inflation
On, Off
On
On
Completion Sound
On, Off
On
On
NIBP Automatic Mode Interval
Off, Cont, 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60,
90,120 ,180 minutes
Off Off
NIBP SYS Upper Alarm Limits
60 to 260 mmHg (Adult), 40 to 130 mmHg (Neo)
8.0 to 34.7 kPa (Adult), 5.3 to 17.3 kPa (Neo)
(10 mmHg / 1.3 kPa steps)
200 mmHg
26.7 kPa
130 mmHg
17.3 kPa
NIBP SYS Lower Alarm Limits
50 to 250 mmHg (Adult), 30 to 120 mmHg (Neo)
6.7 to 33.3 kPa (Adult), 4.0to 16.0 kPa (Neo)
(10 mmHg / 1.3 kPa steps)
70 mmHg
9.3 kPa
50 mmHg
6.7 kPa
NIBP DIA Upper Alarm Limits
40 to 210 mmHg (Adult), 20 to 100 mmHg (Neo)
5.3 to 28.0 kPa (Adult), 2.7 to 13.3 kPa (Adult)
(10 mmHg / 1.3 kPa steps)
160 mmHg
21.3 kPa
100 mmHg
13.3 kPa
NIBP DIA Lower Alarm Limits
30 to 200 mmHg (Adult), 10 to 90 mmHg (Neo)
4.0 to 26.7 kPa (Adult), 1.3 to 12.0 kPa (Neo)
(10 mmHg / 1.3 kPa steps)
30 mmHg
4.0 kPa
10 mmHg
1.3 kPa
NIBP MAP Upper Alarm Limits
50 to 240 mmHg (Adult), 30 to 110 mmHg (Neo)
6.7 to 32.0 kPa (Adult), 4.0 to 14.7 kPa (Neo)
(10 mmHg / 1.3 kPa steps)
180 mmHg
24.0 kPa
110 mmHg
14.6 kPa
NIBP MAP Lower Alarm Limits
40 to 230 mmHg (Adult), 20 to 100 mmHg (Neo)
5.3 to 30.7 kPa (Adult), 2.7 to 13.3 kPa (Neo)
(10 mmHg / 1.3 kPa steps)
40 mmHg
5.3 kPa
20 mmHg
2.7 kPa
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...