35
Technical Alarm
Parameter
Condition
Messages
Deflation time is too long
Deflation speed is too low (C13)
NIBP: Cuff excessive artifact (C13)
Maximum inflating pressure is too low to
calculate patient BP value. (C14)
NIBP: Cuff insufficient pressure
(C14)
Abnormal pulse due to arrhythmia and too
much noise. (C15)
NIBP: Cuff irregular pulses (C15)
Collected pulse’s movement is abnormal.
(C16)
NIBP: Cuff motion artifact (C16)
Measurement time is beyond specified
time. Specified time 160sec for adult and
80sec for neo. (C17)
NIBP: Cuff time-out (C17)
Number of detected pulses is beyond
specified number. Specified number is 160
for both adult and neo. (C18)
NIBP: Cuff time-out, over 160 pulses
(C18)
Cuff pressure is beyond the pressure
specified for the patient safety. Specified
pressure is 300mmHg for adult and
150mmHg for neo. (C19)
NIBP: Cuff pressure failure (C19)
The maximum value of collected pulse is
too low. (C20)
NIBP: Cuff weak pulse (C20)
Cuff size is inadequate to patient. (C21)
NIBP: Check cuff, hose and mode
(C21)
NIBP module error
BPM pressure sensor fault.
Pump operated for ten seconds, however
pressure does not change.
Check the connection of the cuff hose.
NIBP: Internal error (E03)
NIBP module error
NIBP: Internal error (E07)
NIBP module error
NIBP: Internal error (E08)
NIBP module error
Fault detected in accordance with safety
monitoring to BPM IEC standards.
The pressure inside the cuff reaches the
standard pressure,
Standard pressure
Adult: 320mmHg
Neonatal: 157mmHg
NIBP: Internal error (E09)
NIBP module error
NIBP: Internal error (ROM)
NIBP module error
NIBP: Internal error (RAM)
NIBP
NIBP module error
NIBP: Internal error (COM)
ECG
ECG module problem.
ECG: Internal error.
Resp RESP
module
problem.
Resp: Internal error.
IBP signal is not detected.
{label}
: Loss of pulse.
IBP
IBP module problem.
IBP: Internal error.
SpO
2
signal is not detected.
SpO
2
: Loss of pulse.
%SpO
2
SpO
2
module problem.
SpO
2
: Internal error.
Temp
Temperature module problem.
TEMP
{n}
: Internal error.
Capno module have problem.
Capno: Internal error.
Capno
CO
2
sensor error
Capno: Sensor error.
Summary of Contents for BP-S510
Page 10: ...4 This page is intentionally left blank...
Page 22: ...16 This page is intentionally left blank...
Page 38: ...32 This page is intentionally left blank...
Page 68: ...62 This page is intentionally left blank...
Page 80: ...74 This page is intentionally left blank...
Page 104: ...98 This page is intentionally left blank...
Page 108: ...102 This page is intentionally left blank...
Page 114: ...108 This page is intentionally left blank...
Page 122: ...116 This page is intentionally left blank...