35
APPENDIX
BALANCE SPECIFICATIONS
PBL-203
PBL-602(.N)
PBL-2002(.N)
PBL-3101(.N)
Catalog Number
10402-51
10402-52
10402-53
10402-54
Maximum capacity
200g
600g
2000g
3100g
Readability [d]
0.001g
0.01g
0.01g
0.1g
Tare range
-200g
-600g
-2000g
-3100g
Repeatability*
0.002g
0.01g
0.01g
0.1g
Linearity
±0.004g
±0.02g
±0.03g
±0.3g
Stabilization time
2s
2s
2s
2s
Adjustment
External
Display
LCD (with backlit)
NTEP Verified Unit Option
NO
YES
YES
YES
IP rating
IP 43
IP 43
IP 43
IP 43
RS 232
1
1
1
1
Power supply
100 ÷ 240 V AC 50 ÷ 60 Hz / 12 V DC + battery
Battery operating time
(average time)
33h
Operating temperature
+15 to
+30 ˚C
Weighing pan
dimensions
Ø100mm
128x128mm
Packaging dimensions [mm]
330x230x140
Interface
USB type A, USB type B, RS 232
* - Standard deviation
MAITENANCE
Disassembly of weighing pan and other detachable components (the components differ depending
on a balance type – see: UNPACKING AND INSTALLATION section).
Caution:
Cleaning anti-draft chamber while still installed may cause damage to the measuring system
.
13.2.1 CLEANING ABS COMPONETS
To clean dry surfaces and avoid smudging, use clean non-coloring cloths made of cellulose or
cotton. You can use a solution of water and detergent (soap, dishwashing detergent, glass cleaner).
Gently rub the cleaned surface and let it dry. Repeat cleaning process if needed.
In the case when contamination is hard to remove, e.g. adhesive, rubber, resin, polyurethane foam
residues etc., you can use a special cleaning agents based on a mixture of aliphatic hydrocarbons
that do not dissolve plastics. Before using the cleanser for all surfaces, we recommend carrying out
tests. Do not use products containing abrasive substances.