30
9.1.3 PRINTOUT INTERVAL FOR CONTINUOUS TRANSMISSION
Interval for printouts is set in seconds with 0.1[s] accuracy. You can set any interval value
ranging from 0.1 to 3600 seconds.
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<5.1.PC /
5.1.3.Int>
submenu. Message box for entering interval value is displayed.
•
Press
key to confirm. Return to home screen.
9.2 PRINTER
<5.2.Prtr>
menu enables selecting port to which data is send on pressing
key. Content of sent
data is set in
<P6.Prnt>
submenu and described further down this user manual.
9.2.1 PRINTER PORT
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<5.2.Prtr /
5.2.1.Prt>
submenu.
•
Press
key to view available options:
nonE
-
None
rS232
-
RS232 port
USbB
-
USB type B port for connecting a computer.
•
Press
key to confirm. Return to home screen.
PRINTOUTS
<P6.Prnt>
menu enables defining printout templates of:
•
Adjustment report
•
GLP printout
ADJUSTMENT REPORT PRINTOUT
<6.1.CrEP>
menu enables declaring data that is to be printed on adjustment printout.
List of data to be declared:
No.
Name Overview
6.1.1. CtP
Adjustment type
6.1.2. dAt
Adjustment date
6.1.3. tin
Adjustment time
6.1.4. Idb
Balance serial number
6.1.5. CdF
Difference between mass of adjustment weight that was measured during
last adjustment and mass of currently measured adjustment weight.
6.1.6. dSh
Line separating data and signature fields on a printout.
6.1.7. SiG
An area for the signature of a user performing the adjustment.