23
•
Enter the value and press
key to confirm. Message
<SEt Hi>
is displayed for 1
s. Next, a message box for declaring high threshold (Max) is displayed.
•
Enter the value and press
key to confirm. Working mode home screen is
displayed. Threshold value is displayed in the upper part of the display.
Caution:
<Err Lo> message is displayed:
•
If value of entered low threshold (Min) is greater than high threshold value (Max).
<Err Hi> message is displayed:
•
If value of entered high threshold (Max) is greater than the maximum capacity.
7.9 . PERCENT WEIGHING AGAINST REFERENCE SAMPLE MASS
Working mode enables comparison of a measured sample with the reference mass. The result is
expressed in
[%].
Reference mass can be determined by weighing or entered to balance's memory.
7.9.1 SETTINGS
•
Accessibility of
<dEu>
(% Weighing) mode is declared within
<P3.4.1Acc>,
procedure
found within section 7.2
•
Method of sending data to peripheral device is declared within
<34.2.Snn>
, procedure found
within section 7.3.
•
Method of setting automatic printout time for sending data to peripheral device is declared
within
<3.4.3.Snn>
, procedure found within section 7.4.
•
Method of setting lo threshold value for automatic operation is declared within
<3.4.4.Lo>
,
procedure found within section 7.5.
7.9.2 OPERATION MODE
Parameter allowing you to select method of determination of sample piece mass.
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<3.4.dEu
/ 3.4.2.UUt>
submenu.
•
Press
key, parameter values are displayed successively one by one:
S_S
Select to set reference sample mass by determining the mass value.
Suu
Select to set reference sample mass by entering the mass value.
•
Enter respective value and press
key to confirm, then continue weighing.