Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<P2.rEAd
/ 2.5.tArA>
submenu.
•
Press
key to view available options:
no
-
Basic tare mode. Set (selected) tare value is overwritten on
entering new tare value.
tArF
- Last tare value is stored in balance's memory. Tare value is automatically
displayed on restarting the balance.
AtAr
- Tare value is saved after the power supply is disconnected.
EAcH
- Automatic taring of each approved measurement.
•
Press
key to confirm. Return to home screen.
LAST DIGIT
Function enables displaying the last digit of decimal place for a weighing result the measurement is
carried out with lesser accuracy.
Procedure
:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<P2.rEAd
/ 2.6.LdiG>
submenu.
•
Press
key to view available options:
ALAS
-
All digits visible
nEur
-
Last digit is not displayed
uuSt
-
Last digit is displayed only for a stable weighing result
•
Press
key to confirm. Return to home screen.
ADJUSTMENT
In order to ensure the highest weighing accuracy, it is recommended to periodically introduce a
corrective factor of indications to balance memory, the said factor must be referred to a reference mass.
It is a balance adjustment.
Adjustment has to be carried out:
•
Prior to weighing
•
If long breaks between following measuring series occur
•
If the ambient temperature has changed dynamically
•
If the balance's place of use has changed
Types of adjustment:
•
External adjustment
<1.1.CA-E>
carried out with external weight of declared mass which
cannot be modified.
•
User adjustment
<1.2.CA-u>
carried out with any external weight of mass equal or greater than
30% of maximum capacity.
EXTERNAL ADJUSTMENT
External adjustment is carried out using external weight of class F1.