19
AUTOMATIC PRINTOUT TIME INTERVAL
Parameter enabling you to set frequency of automatic printout. Printout interval is set in minutes with
1 [min] accuracy within 1 [min] - 1440 [min] range.
Utilize
and
to set interval time in minutes within 1 [min] – 1440 [min]. To confirm value
press
.
Caution:
Option not available for Peak Hold working mode.
LO THRESHOLD
<Lo>
parameter allows you to configure the function of automatic operation. In order to save the
next measurement, before carrying it out the mass indication must get below the set net value of Lo
threshold.