16
•
On unloading the weighing pan, submenu
<1.1.CA-E>
is displayed.
USER ADJUSTMENT
User adjustment is carried out using external weight of class F1.
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<P1.CAL /
1.2.CA-u>
submenu. Message box for entering mass of a weight used for user
adjustment is displayed. Mass of the weight has to be equal or greater than 30% of
maximum capacity.
•
On entering and confirming the weight mass, message
<UnLoAd>
(unload the weighing
pan) is displayed.
•
On unloading the weighing pan, press
key.
•
Balance starts determining mass of an empty weighing pan. The process is signaled by
dashes
<- >
. Next, message
<Load>
(put weight) is displayed along with declared mass
value to be put onto weighing pan; e.g..
100.000g
.
•
Put the weight of a given mass and press
key.
•
Balance starts determining mass of a weight. The process is signaled by dashes
<- >
.
Next, message
<UnLoad>
(remove weight) is displayed.
•
On unloading the weighing pan submenu
<1.2.CA-u>
is displayed.
ADJUSTMENT REPORT
Adjustment report is automatically printed on a printer connected to the balance at the end of each
adjustment. Report content is to be declared in
<P6.1.CrEP>
menu and described further down
this user manual.
WORKING MODES
The balance features the following working modes: