Preparing the microscope for use

Carefully remove the items from its packaging and place them on a flat, firm surface. Please 
do not expose the microscope to direct sun light, high temperatures, damp, dust or acute 
shake. Make sure the table or surface is flat and horizontal 

When moving the microscope, use the left hand to hold the transport handle (A) and 
hold the base of the microscope with the right hand

Caution!

 Hold the microscope at the top of the stand arm when it is moved

Insert the power cord into the back of the microscope and use 
the cable storage system (CSS) to store excessive cable while in 
use or to store the cable after use (B)

Caution!

 If the bacterial solution or water splatters over the 

stage, objective or head, pull out the power cord immediately 
and dry the microscope

Assembling steps 

Mounting the objectives

1. 

Rotate the coarse focusing knob to lower the stage to its 
lowest position

2. 

Install the objectives into the objective nosepiece from 
the lowest magnification to the highest in a clockwise direction from the rear side of the microscope. When 
operating, start using the low magnification objective (4X or 10X) to search for specimen and focus, then 
continue with high magnification objective to observe

The microscope head

The standard MSU-400 series configuration is supplied with the head assembled. However, if your order contains a 
fluorescence attachment (model 78904-13) then this should be mounted first; see pages 14-16 for supplementary 
manual.

Placing the eyepieces 

The standard MSU-400 series configuration is supplied with the eyepieces assembled. However, if your order contains 
additional eyepieces, please follow these steps to mount/replace them:

1. 

Unscrew the current eyepiece from the eyepiece tube with a screwdriver (see pictures A and B)

Note:

 The eyepiece tube with the diopter adjustment has three screws. In order to remove the eyepiece you need to 

unscrew the one aligned with “0” (see picture C) 

2. 

Insert the eyepiece into the eyepiece tube

3. 

Lock the eyepieces with a screwdriver

8

Summary of Contents for 78904-02

Page 1: ...Installation and Operating Instructions MSU 400 Series Compound Microscopes Models 78904 02 04 07 10 13 78904 02 78904 07 78904 13...

Page 2: ..._______________________________________________________ 11 Use of the S100x oil immersion objective_________________________________________________________ 12 Illumination____________________________...

Page 3: ...comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use and are undamaged and...

Page 4: ...sed disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination stateme...

Page 5: ...e force when turning the knobs Make sure that the microscope system can dissipate its heat fire hazard Keep the microscope away from walls and obstructions for at least approximately 15 cm Never turn...

Page 6: ...and S100x objectives are equipped with a spring mount to prevent damage to the front lens and the slide Model Type Objectives Working distance Stand Illumination Power Weight 78904 02 Binocular 4x 10...

Page 7: ...er adjustment B Trinocular tube K Slide for polarization filter C Microscope head L Nosepiece D Transport handle M Objectives E Height adjustment condenser N Stage with X Y mechanical stage F Slide pr...

Page 8: ...2 Install the objectives into the objective nosepiece from the lowest magnification to the highest in a clockwise direction from the rear side of the microscope When operating start using the low magn...

Page 9: ...illumination For optimal contrast and resolution one should follow the below procedure 1 Place a specimen on the object stage and focus using the 4x objective with a fully opened iris diaphragm 2 Tur...

Page 10: ...e field of view see image below This distance can be set by either pulling the tubes towards each other or pulling them away from each other This distance is different for each observer and thus shoul...

Page 11: ...aphragm gradually 7 The condenser is centered correctly if the image remains in the center when you open the field diaphragm and inscribed to the field of view The field K hler diaphragm see A in pict...

Page 12: ...the front lens doesn t touch the oil any longer Clean the front lens of the S100x objective 12 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a dr...

Page 13: ...te the 10 infinity phase contrast objective into the field 10 to match the objective Take the eyepiece out of the tube and insert centering telesc telescope the dark and bright ring images should coin...

Page 14: ...reen F Fluorescence illuminator G Turret for LEDs Installation steps Take the microscope with fluorescence attachment fluorescence illuminator and head from its packaging Place the head on the fluores...

Page 15: ...escence filters Blue Green Violet and Ultraviolet mounted in the turret fig 1 D and are indicated as B G V and UV on the turret There is also a brightfield option and it is indicated as 0 on the turre...

Page 16: ...steps 1 Use objective 10x 2 Push in the field diaphragm switch the one in the front fig 2 A a light spot will be visible in the field of view 3 Adjust the centering if the light spot deviates from th...

Page 17: ...Cleaning the optics When the eyepiece lens or front lens of the 10x or S40x objective are dirty they can be cleaned by wiping a piece of lens paper over the surface circular movements When this does...

Page 18: ...roscopes come with a five 5 year factory warranty against manufacturing defects covering labor and parts Warranty does not apply for consumable accessories and parts such as bulbs batteries fuses cord...

Page 19: ..._ ___________________________________________________________________________________________________ __________________________________________________________________________________________________...

Page 20: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 400_Series_Compound Microscopes_Manual...

Reviews: