PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
80
9.4.43
:OUTPut<n>:DATA<m>:CPOint
Command :OUTPut<n>:DATA<m>:CPOint
Description Sets the pattern cross point (%).
Syntax
:OUTPut<n>:DATA<m>:CPOint<wsp><value|MIN|MAX|DEF>
Parameters
•
value: The desired new set point
•
MIN: Sets to the minimum value
•
MAX: Sets to the maximum value
•
DEF: Restores the default value
Response N/A
Example
:OUTPut<n>:DATA<m>:CPOint 35
9.4.44
:OUTPut<n>:DATA<m>:OUTPut?
Command :OUTPut<n>:DATA<m>:OUTPut?
Description Queries the state of the pattern generator output.
Syntax
:OUTPut<n>:DATA<m>:OUTPut?[<wsp><DEF|SET|INFO>]
Parameters
•
DEF: Queries the default power-on value
•
SET: Queries the current SET value
•
INFO: Provides a description of what the return values represent
By default, the command returns the current SET value.
Response The response is an integer indicating the state of the pattern generator
output.
Example
:OUTPut3:DATA2:OUTPut?
-> 1
:OUTPut3:DATA2:OUTPut? INFO
->0:OFF