9.4.1
:CALCulate<n>:DATA<m>:EALarm?
Command :CALCulate<n>:DATA<m>:EALarm?
Description Queries the ED error measurement.
Syntax
:CALCulate<n>:DATA<m>:EALarm?<wsp>
<STATE|DATA|LOCK|ERROR|COUNt|BITS|BER|FULL>
Parameters
•
STATE: Queries the state of the ED measurement
•
DATA: Queries the presence of the data pattern
•
LOCK: Queries the frequency lock status
•
COUNt|ERROR: Queries the total error count
•
BITS: Queries the total number of bits
•
BER: Queries the bit error rate
•
FULL: Queries the information specified by all the parameters
By default, the command queries the total error count.
Response The response is the requested value, or a comma separated array.
When
FULL
is requested, the format is
"
<STATE>,<DATA>,<LOCK>,<COUNt|ERROR>,<BITS>,<BER>
".
Example
:CALCulate3:DATA2:EALarm? FULL
1->1,1,1,5.11,7.03,6.07
9.4.2
:CALCulate<n>:DATA<m>:EALarm:ELAPsed?
Command :CALCulate<n>:DATA<m>:EALarm:ELAPsed?
Description Queries the running time for the ED measurement.
Syntax
:CALCulate<n>:DATA<m>:EALarm:ELAPsed?
Parameters None
Response The response is the number of seconds the ED measurement has
been running.
Example
:CALCulate3:DATA2:EALarm:ELAPsed?
->32.568