PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
76
9.4.36
:OUTPut<n>:CLOCk:FREQuency:STanDard?
Command :OUTPut<n>:CLOCk:FREQuency:STanDard?
Description Queries the current synthesizer frequency standard settings.
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard?[<wsp><MIN|MAX|
DEF|SET|LIST|ALL|STEP|UNIT>]
Parameters
•
MIN: Queries the minimum configurable value
•
MAX: Queries the maximum configurable value
•
DEF: Queries the default power-on value
•
SET: Queries the current SET value
•
LIST: Queries the list of discrete options for configurable values (if
available)
•
ALL: Queries all the values specified by MIN, MAX, DEF, SET, and
LIST
•
STEP: Queries the minimum resolution of the frequency value
•
UNIT: Queries the default unit for the frequency value
By default, this command queries the current SET value.
Response The response is a single value or a comma separated string of values.
For
ALL
, the format of the response is
"
<MIN>,<MAX>,<DEF>,<SET>,<LIST>
".
If the frequency is set to an arbitrary value, the
SET
value in the
response is
NAN
.
Example
STANDARD MODE:
:OUTP3:CLOC:FREQ:STD?
-> 14.5
ARBITRARY MODE:
:OUTP3:CLOC:FREQ:STD?
-> NAN