![Coherent Solutions BERTPXIe-1003 User Manual Download Page 58](http://html1.mh-extra.com/html/coherent-solutions/bertpxie-1003/bertpxie-1003_user-manual_2634920058.webp)
PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
57
9.4.3
:CALCulate<n>:DATA<m>:EDIAGram?
Command :CALCulate<n>:DATA<m>:EDIAGram?
Description Queries the eye scan measurement data.
This command can only be run after
the
:CALCulate<n>:DATA<m>:EDIagram:STATus?
command
has returned
COMPLETE
.
Syntax
:CALCulate<n>:DATA<m>:EDIAGram?
Parameters None
Response The response is the eye scan data points in a comma separated array.
The format of the data is
"
<number_of_rows>,<number_of_columns>,<points[0][0]
>, .. ,<points[number_of_columns-
1][number_of_columns-1]>
"
Note: The number of rows and columns depends on your eye scan
settings.
Example
:CALCulate3:DATA2:EDIAGram?
-> "63,63,<array of 63*63 points>"