![Coherent Solutions BERTPXIe-1003 User Manual Download Page 78](http://html1.mh-extra.com/html/coherent-solutions/bertpxie-1003/bertpxie-1003_user-manual_2634920078.webp)
PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
77
9.4.37
:OUTPut<n>:CLOCk:FREQuency:STanDard
Command :OUTPut<n>:CLOCk:FREQuency:STanDard
Description Sets the current synthesizer frequency to a standard value.
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard<wsp>
<value|MIN|MAX|DEF>
Parameters
•
value: The desired new value
•
MIN: Sets to the minimum value
•
MAX: Sets to the maximum value
•
DEF: Restores the default value
Response N/A
Example
:OUTP3:CLOC:FREQ:STD 14.5
9.4.38
:OUTPut<n>:CLOCk:POWer?
Command :OUTPut<n>:CLOCk:POWer?
Description Queries the power of the RF output clock
Syntax
:OUTPut<n>:CLOCk:POWer?
[<wsp><MIN|MAX|DEF|SET|LIST|ALL|STEP|UNIT>]
Parameters
•
MIN: Queries the minimum configurable value
•
MAX: Queries the maximum configurable value
•
DEF: Queries the default power-on value
•
SET: Queries the current SET value
•
LIST: Queries the list of discrete options for configurable values (if
available)
•
ALL: Queries all the values specified by MIN, MAX, DEF, SET, and
LIST
•
STEP: Queries the minimum resolution of the power value
•
UNIT: Queries the default unit for the power value
By default, the command queries the current SET value.
Response The response is a single value or a comma separated string of values.
For
ALL
, the format of the response is
"
<MIN>,<MAX>,<DEF>,<SET>,<LIST>
".
Example
:OUTPut3:CLOCk:POWer?
-> 0