Maintenance and Support
161
RetCam 3 and RetCam Shuttle User Manual
PN 18-000393 Rev. A
Figure 154
Removing the Fuse Holder
3.
Remove the blown fuse from the fuse holder (
) and replace it
with a new fuse of same type and rating. Reinsert the fuse holder.
Figure 155
Blown Fuse Removed
For continued protection against risk of fire, replace only with same type and
rating fuse.
Replacement fuses are available from Clarity Technical Services.
RetCam Shuttle Illumination Lamp Replacement
Appropriately shutdown and unplug the unit to allow the bulb to cool before
replacing it.
NOTE
Replacement bulbs are available from Clarity Technical Services.
Cover open
Gently pry at edges
Finish removal with fingers