160
Chapter 12
Maintenance and Support
RetCam 3 and RetCam Shuttle User Manual
PN 18-000393 Rev. A
RetCam Shuttle Maintenance
This section describes the specific recommended process for fuse replacement and
illumination lamp replacement for the RetCam Shuttle system. These activities are
appropriate under the following conditions:
•
Blown fuse:
Replace system fuses.
•
Burned out bulb:
Replace the illumination bulb.
RetCam Shuttle Fuse Replacement
Do not replace the fuse until the system is completely shut down. Locate the
power entry module on the lower right side of the rear of the cart (
). Make sure to remove the power cord before proceeding any
further.
Figure 152
Power Entry Module
To replace the RetCam Shuttle fuse:
1.
Using a flat blade screwdriver, gently pry open the cover of the fuse holder at the top of
the power entry module (see
).
Figure 153
Opening the Fuse Holder Cover
2.
Remove the fuse holder (see