Maintenance and
service
Learn the information about CounterAct maintenance and service.
Warning! If not avoided, the following could result in death or serious injury.
•
SHOCK HAZARD! Disconnect the product from AC before installing, moving, servicing, cleaning,
removing components, or opening any enclosure.
•
ELECTRICAL and BURN HAZARD! Use caution when accessing internal components.
•
PERSONAL INJURY HAZARD! Eye damage may result from directly viewing the light produced by
the lamp in this product. Always turn off the lamp before performing maintenance or service.
•
PERSONAL INJURY HAZARD! Use caution when servicing the lamp to avoid the risk of personal
injury.
Caution! If not avoided, the following could result in minor or moderate injury.
•
Always connect the ground or earth first to reduce shock hazard.
•
This Christie product must be serviced by Christie qualified technicians.
•
HOT SURFACE HAZARD! Do not touch the lamp module during operation.
•
HOT SURFACE HAZARD! Allow lamp sufficient time to cool before handling.
•
SHOCK HAZARD! Always turn off power before performing lamp module maintenance.
•
FIRE HAZARD! Keep all flammable items, such as paper or cloth, away from the lamp module.
•
Use protective eye wear and gloves when handling the lamp module.
•
Do not drop, hit, use force, or scratch the lamp module.
•
Do not disassemble or modify the lamp module.
•
Use only mercury-free, excimer lamps (UXFL70-222B4-UIA) sold by Christie (P/N: 174-006107-
XX/003-121997-XX).
Outside of lamp replacements, this product is not serviceable in the field and must be sent back to
Christie if servicing is required.
Replacing the lamps
Replace the lamps when instructed by the CounterAct application.
1. Disable power to the system.
2. Remove the disinfecting module you want to replace the lamps in (on page 18).
3. Unplug the connector between the lighting module and the integrator board.
CounterAct Installation and Setup Guide
45
020-103456-02 Rev. 1 (12-2021)
Copyright
©
2021 Christie Digital Systems USA, Inc. All rights reserved.