Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
8 Maintenance and service
60 / 84
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
Translation of the original operating manual
• Separate the magnetic coupling from the rotor shaft. To do so, unscrew
the four slotted screws that join the magnetic coupling with the rotor
shaft.
7
Slotted screws
8
Magnetic coupling,
separate
Fig. 49: Rotor shaft and upper ball bearing
• Remove the lower ball bearing.
Fig. 50: Removal of the lower ball bearing
• Clean the rotor shaft and magnetic coupling.
• Insert a new lower ball bearing.
• Screw the magnetic coupling and rotor shaft together.