Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
12 Glossary
80 / 84
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
Translation of the original operating manual
Safety pressure
Since the vacuum has a dominating influence on the product temperature,
Martin Christ Gefriertrocknungsanlagen GmbH has integrated a so-called
safety pressure function into their units in order to protect the products. If
the pressure inside the rotor chamber rises too quickly
– above the safety
limit
– the energy supply of the unit will be interrupted so that the
evaporation process slows down.
Fig. 55: Diagram describing the safety pressure function