Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
8 Maintenance and service
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
59 / 84
Translation of the original operating manual
• Clean the lid with a disinfectant or soap water.
• Wipe the rotor chamber clean and clean it with soap water or a suitable
disinfectant.
Cleaning the rotor shaft and magnetic coupling
CAUTION
Ensure that no cleaning agent or disinfectant gets into the ball bearings of
the rotor shaft as it may wash out the ball bearing grease.
This can lead to irreparable damage.
7
Rotor shaft with
magnetic coupling
8
Lower ball bearing in
the rotor shaft
Fig. 47: Rotor shaft and lower ball bearing
• Remove the upper ball bearing from the rotor shaft.
9
Upper ball bearing
Fig. 48: Rotor shaft and upper ball bearing