7 Malfunctions and error correction
110 / 148
Version 04/2014, Rev. 2.13 of 01/12/2021
• sb
Translation of the original operating manual
Pos : 218 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020 Kein ausr eichendes Vakuum @ 25\m od_1404984060039_68.docx @ 184167 @ 3 @ 1
→
7.1.2
Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is
frozen.
Pos : 219 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 220 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0010 Klei nflansc hv erbi ndungen @ 25\m od_1404984061086_68. docx @ 184181 @ 4 @ 1
→
7.1.2.1
Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 73: Small flange and centring ring
Fig. 74: Small flange with centring ring
and small flange
Fig. 75: Attaching the clamping ring
Fig. 76: Tightened clamping ring